Z-0118-2019Class II
AS-OBGYN Information System version 7.824.x
As Software, LLC / initiated 2018-09-20 / Terminated / root cause: design
AS Software, Inc. began a recall of AS-OBGYN Information System version 7.824.x on . FDA classified it as Class II and gives the reason as "Any change made to current ultrasound machine software or hardware upgrades, adding or modifying custom fields, or adding a new Ultrasound Machine, must be tested prior to the use in production environment to ensure...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0118-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- As Software, LLC
- Product
- AS-OBGYN Information System version 7.824.x
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Any change made to current ultrasound machine software or hardware upgrades, adding or modifying custom fields, or adding a new Ultrasound Machine, must be tested prior to the use in production environment to ensure that ultrasound measurements are crossing over correctly to the AS system
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- AS Software issued letter by email on 9/20/18 to notify customers that any change made to their ultrasound machine such as software or hardware upgrades, adding or modifying custom fields, or adding a new Ultrasound Machine, are tested prior to the use in the production environment. Medical Device Correction Return Response form to be signed and returned. The AS Software Support Department can be reached Monday through Friday, 9:00 AM to 6:00 PM EST via phone: 201-541-1900 option 2 or e-mail: [email protected]
- Quantity in commerce
- 52 systems
- Distribution
- US Nationwide.
- Product codes and lots
- Software version 7.824.x
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0118-2019 | Class II | Terminated | Voluntary: Firm initiated |