Z-0249-2020Class II

EndoWrist Suction Irrigator, Single Use Instrument. Part Number: 480299-04. An accessory to the EndoWrist...

Intuitive Surgical, Inc. / initiated 2018-06-04 / Open, Classified

Intuitive Surgical, Inc. began a recall of EndoWrist Suction Irrigator, Single Use Instrument. Part Number: 480299-04. An accessory to the EndoWrist Surgical System, used in endoscopic surgery on . FDA classified it as Class II and gives the reason as "The firm has identified the potential for a silicone particle to be generated within the Suction Irrigator device. If generated, a particle may be introduced into the patient with irrigation". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-0249-2020
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Intuitive Surgical, Inc.
Product
EndoWrist Suction Irrigator, Single Use Instrument. Part Number: 480299-04. An accessory to the EndoWrist Surgical System, used in endoscopic surgery.
Product code
NAY System, Surgical, Computer Controlled Instrument
Reason for recall
The firm has identified the potential for a silicone particle to be generated within the Suction Irrigator device. If generated, a particle may be introduced into the patient with irrigation.
FDA root cause
Process change control
Action
On 06/04/2018, Medical Device Recall notices (Field Safety Notice Urgent Medical Device Correction- ISIFA2018-09-C) were emailed to U.S. consignees, and on 06/07/2018 notices were mailed via FedEx. Additionally on 06/08/2018 two (2) distributors received the notices via email and on 06/15/2018 two (2) EMEIA consignees received the notices by mail. The firm posted information regarding the voluntary field recall on their US Customer Portal on 06/04/2018. Customers were asked to: 1. Stop-use of the affected products for intra-cardiac procedures until further notice from the Recalling Firm. 2. Ensure surgeons and patient side assistants using the products have read and understand the content of the notification letter. 3. Complete the attached Acknowledgement Form and return it via fax or email to U.S. FAX +1(408)523-0619, or Scan and Email [email protected] or for further questions contact Customer Service at: -North and South America: 800-876-1310 Option 3 (4 am to 5 pm PST) -Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 am to 6 pm CET) 4. Please retain a copy of the Customer Notification and the Acknowledgement form on file.
Quantity in commerce
3108 units (518 boxes of 6 units)
Distribution
Worldwide distribution. US nationwide, Australia, Italy, Switzerland, and United Kingdom.
Product codes and lots
Lots: T10180226 and onward; Part Numbers: 480299-04 (UDI 00886874114810)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0249-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0249-2020Class IIOngoingVoluntary: Firm initiated