Z-0585-2019Class II

B20i Patient Monitor and B20i V2 Patient Monitor. The device is a portable multiparameter unit to be used for...

GE Healthcare / initiated 2018-09-11 / Open, Classified / root cause: software

GE Healthcare, LLC began a recall of B20i Patient Monitor and B20i V2 Patient Monitor. The device is a portable multiparameter unit to be used for monitoring and recording of, and to generate... on . FDA classified it as Class II and gives the reason as "Patient monitors may restart due to network overload caused by network configuration". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0585-2019
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
GE Healthcare
Product
B20i Patient Monitor and B20i V2 Patient Monitor. The device is a portable multiparameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport.
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
Patient monitors may restart due to network overload caused by network configuration.
Root cause tag
software
FDA root cause
Software design
Action
The recalling firm, GE Healthcare, issued "URGENT MEDICAL DEVICE CORRECTION" letters dated 9/11/2018 via trackable mail on 9/11/2018 to its customers. The letters described the product, problem and actions to be taken. The letters informed the customer of the issue and providing safety instructions for mitigation of the issue if a prolonged network overload occurs and the Bx0 and/or B1x5 patient monitor on the network does not resume normal functionality in approximately 90 seconds. The customers were instructed to follow the safety instructions. GE Healthcare will provide revised network installation and user instructions documentation at no cost to you in a future mailing. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Quantity in commerce
6,458 patient monitors
Distribution
Worldwide Distribution: US (nationwide) including Puerto Rico; and countries of: Algeria, Argentina, Australia, Austria, Azerbaijan, Bahamas, Bangladesh, Belgium, Bolivia, Bosnia-Herzegovina, Brazil, Brunei, Bulgaria, Burma (Myanmar), Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Ethiopia, Finland, France, Gabon, Georgia, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Ivory Coast, Jamaica, Jordan, Kampuchea, Kazakstan, Kenya, Korea, Kuwait, Lebanon, Lithuania, Macedonia, Malaysia, Mexico, Moldova, Morocco, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Senegal, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Viet Nam, and Zambia.
Product codes and lots
All monitors

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0585-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0583-2019Class IIOngoingVoluntary: Firm initiated
Z-0584-2019Class IIOngoingVoluntary: Firm initiated
Z-0585-2019Class IIOngoingVoluntary: Firm initiated
Z-0586-2019Class IIOngoingVoluntary: Firm initiated
Z-0587-2019Class IIOngoingVoluntary: Firm initiated