Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0170-2019Class II

Mindray BeneVision Distributed Monitoring System (DMS), Part No. 115-034413-00

Datascope Corp. / initiated 2018-08-30 / Terminated / root cause: software

Mindray DS USA, Inc. dba Mindray North America began a recall of Mindray BeneVision Distributed Monitoring System (DMS), Part No. 115-034413-00 on . FDA classified it as Class II and gives the reason as "The firm has identified performance anomalies with the device caused when more than 64 TD60 telemetry transmitters have been connected to a central station. The anomaly causes some ECG waveforms to appear as a flat line...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0170-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Datascope Corp.
Product
Mindray BeneVision Distributed Monitoring System (DMS), Part No. 115-034413-00
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
The firm has identified performance anomalies with the device caused when more than 64 TD60 telemetry transmitters have been connected to a central station. The anomaly causes some ECG waveforms to appear as a flat line at the centralized viewing station and will manifest if more than 64 TD60 transmitters have been connected to a given central station server. Each central station supports only 32 transmitters at a time. For the issue to manifest, the BeneVision DMS needs to be configured to use flexible device assignment, have more than 64 TD60 transmitters, and the workflow must allow 64 TD60s to move around between multiple central stations in the same care area.
Root cause tag
software
FDA root cause
Software design
Action
The firm notified affected consignees on August 30, 2018, via Urgent Medical Device Recall letter. The letter advised customers of the software issue when more than 64 telemetry transmitters have been connected to a central station. Customers were advised to use the system normally, but informed that a Mindray Service Representative would be in contact to arrange for a system software update to be performed at the customer site. Mindrays Technical Support team may be reached at (877) 913-9663 (Option 1), Monday through Friday, 8:30am - 5:30pm ET should you have technical questions. Customers were also asked to complete and return an attached Acknowledgement and Receipt Form.
Quantity in commerce
5
Distribution
Distributed to accounts in CA, CO, FL, MA, NY, PA, PR, TX, and WV.
Product codes and lots
This potential issue affects BeneVision DMS with software version 03.01.00 to 03.03.00

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0170-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0170-2019Class IITerminatedVoluntary: Firm initiated