Mindray BeneVision Distributed Monitoring System (DMS), Part No. 115-034413-00
Mindray DS USA, Inc. dba Mindray North America began a recall of Mindray BeneVision Distributed Monitoring System (DMS), Part No. 115-034413-00 on . FDA classified it as Class II and gives the reason as "The firm has identified performance anomalies with the device caused when more than 64 TD60 telemetry transmitters have been connected to a central station. The anomaly causes some ECG waveforms to appear as a flat line...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0170-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Datascope Corp.
- Product
- Mindray BeneVision Distributed Monitoring System (DMS), Part No. 115-034413-00
- Product code
- MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Reason for recall
- The firm has identified performance anomalies with the device caused when more than 64 TD60 telemetry transmitters have been connected to a central station. The anomaly causes some ECG waveforms to appear as a flat line at the centralized viewing station and will manifest if more than 64 TD60 transmitters have been connected to a given central station server. Each central station supports only 32 transmitters at a time. For the issue to manifest, the BeneVision DMS needs to be configured to use flexible device assignment, have more than 64 TD60 transmitters, and the workflow must allow 64 TD60s to move around between multiple central stations in the same care area.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm notified affected consignees on August 30, 2018, via Urgent Medical Device Recall letter. The letter advised customers of the software issue when more than 64 telemetry transmitters have been connected to a central station. Customers were advised to use the system normally, but informed that a Mindray Service Representative would be in contact to arrange for a system software update to be performed at the customer site. Mindrays Technical Support team may be reached at (877) 913-9663 (Option 1), Monday through Friday, 8:30am - 5:30pm ET should you have technical questions. Customers were also asked to complete and return an attached Acknowledgement and Receipt Form.
- Quantity in commerce
- 5
- Distribution
- Distributed to accounts in CA, CO, FL, MA, NY, PA, PR, TX, and WV.
- Product codes and lots
- This potential issue affects BeneVision DMS with software version 03.01.00 to 03.03.00
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0170-2019 | Class II | Terminated | Voluntary: Firm initiated |