Tritanium TL Steerable Inserter, Cat. No. 48930000, UDI 07613327381900

Howmedica Osteonics Corp / initiated 2018-09-12 / Terminated

Howmedica Osteonics Corp. began a recall of Tritanium TL Steerable Inserter, Cat. No. 48930000, UDI 07613327381900 Product Usage: The TL Steerable Inserter is a Class 1 instrument which is provided in a... on . FDA classified it as Class II and gives the reason as "The firm received complaints related to the locking mechanism on the steerable inserter not functioning as intended. The firm confirmed that the issue is related to lack of lubrication of the instrument. There are two...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0207-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Tritanium TL Steerable Inserter, Cat. No. 48930000, UDI 07613327381900 Product Usage: The TL Steerable Inserter is a Class 1 instrument which is provided in a non-sterile condition. The inserter is designed to enter the disc space, with a fixed engagement between the inserter and the cage and then rotate for desired placement. The three components of the TL Steerable Inserter System include: 1. TL Steerable Inserter 2. TL Steerable Inserter Inner Shaft 3. TL Steerable Inserter Driver
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
The firm received complaints related to the locking mechanism on the steerable inserter not functioning as intended. The firm confirmed that the issue is related to lack of lubrication of the instrument. There are two potential hazards associated with the implant pivoting unexpectedly during implant insertion into the disc space. First, if the implant is initially inserted into the disc space via the impaction method and the implant pivots unexpectedly, the implant potentially may unintentionally contact sensitive anatomy. Additionally, the user may potentially experience difficulty positioning the implant within the disc space due to the lack of rigidity. A third potential hazard is associated with the inserter not unlocking at the articulating pivot head during implant positioning within the disc space. Once the implant is in the prepared disc space and contacts the ventral annulus, the surgeon unlocks the articulating pivot head and positions the implant perpendicular to the spinous process. The surgeon may then elect to lock/unlock the articulating pivot head for desired final positioning of the implant. If the inserter does not unlock during these final implant positioning steps, the user may potentially implant the cage in a suboptimal trajectory/placement.
FDA root cause
Under Investigation by firm
Action
On September 12, 2018, the firm notified its sales personnel of the recall via a Product Correction letter. The letter identified the affected product, problem and actions to be taken. Applicable Stryker personnel that are involved in the maintenance or preparation of the Tritanium TL Inserter have been identified to receive notification of this correction to assure that the instrument is lubricated adequately.
Quantity in commerce
58
Distribution
US Nationwide Distribution.
Product codes and lots
All lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0207-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0207-2019Class IITerminatedVoluntary: Firm initiated