Z-0491-2019Class II
BD Vacutainer Urine Collection Cups, Cat. No. 364941
Becton Dickinson & Co / initiated 2018-08-30 / Terminated
Becton Dickinson & Company began a recall of BD Vacutainer Urine Collection Cups, Cat. No. 364941 on . FDA classified it as Class II and gives the reason as "BD Vacutainer Urine Collection Cups may contribute to a slight elevation of Red Blood Cell (RBC) counts when used with the Sysmex UF Series Analyzers. BD recommends following Good Clinical Practice (GCP) by using...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0491-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- BD Vacutainer Urine Collection Cups, Cat. No. 364941
- Product code
- JTW System, Transport, Aerobic
- Reason for recall
- BD Vacutainer Urine Collection Cups may contribute to a slight elevation of Red Blood Cell (RBC) counts when used with the Sysmex UF Series Analyzers. BD recommends following Good Clinical Practice (GCP) by using microscopy to confirm an elevated RBC Count in a Urine sample as per CLSI GP16-3A. 29(4) and CAP Checklist on Urinalysis. Product will continue to ship to not impact critical patient care.
- FDA root cause
- Under Investigation by firm
- Action
- On August 30, 2018, BD notified Sysmex via an emailed Urgent Medical Device Product Safety Alert letter. Sysmex was informed that BD had confirmed the product issue with BD Vacutainer Urine Collection Cups were used with Sysmex UF Series Analyzers. Sysmex was instructed to take the following actions: - Notify Sysmex customers of the potential for elevation of RBC counts - Complete the customer Resopnse form and email it to BD at [email protected] to confirm receipt of the Safety Alert. Product will continue to ship so as to not impact critical patient care. BD recommends following Good Clinical Practice (GCP) by performing microscopy to confirm an elevated RBC Count in a urine sample as per CLSI GP16-3A.29(4) and CAP Checklist on Urinalysis.
- Quantity in commerce
- Total: 29,982,150 BD cups (US); 97,800,900 BD cups (OUS)
- Distribution
- Illinois
- Product codes and lots
- All lots within expiry
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0491-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0492-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0493-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0494-2019 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
