KIT URIN CUP PLC 16X100 10.0 UA YEL, Cat. No. 364981: Urine Collection Kit for collection and transport of...
Becton Dickinson & Company began a recall of KIT URIN CUP PLC 16X100 10.0 UA YEL, Cat. No. 364981: Urine Collection Kit for collection and transport of urine specimens for urinalysis on . FDA classified it as Class II and gives the reason as "BD Vacutainer Urine Collection Cups may contribute to a slight elevation of Red Blood Cell (RBC) counts when used with the Sysmex UF Series Analyzers. BD recommends following Good Clinical Practice (GCP) by using...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0493-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- KIT URIN CUP PLC 16X100 10.0 UA YEL, Cat. No. 364981: Urine Collection Kit for collection and transport of urine specimens for urinalysis
- Product code
- JTW System, Transport, Aerobic
- Reason for recall
- BD Vacutainer Urine Collection Cups may contribute to a slight elevation of Red Blood Cell (RBC) counts when used with the Sysmex UF Series Analyzers. BD recommends following Good Clinical Practice (GCP) by using microscopy to confirm an elevated RBC Count in a Urine sample as per CLSI GP16-3A. 29(4) and CAP Checklist on Urinalysis. Product will continue to ship to not impact critical patient care.
- FDA root cause
- Under Investigation by firm
- Action
- On August 30, 2018, BD notified Sysmex via an emailed Urgent Medical Device Product Safety Alert letter. Sysmex was informed that BD had confirmed the product issue with BD Vacutainer Urine Collection Cups were used with Sysmex UF Series Analyzers. Sysmex was instructed to take the following actions: - Notify Sysmex customers of the potential for elevation of RBC counts - Complete the customer Resopnse form and email it to BD at [email protected] to confirm receipt of the Safety Alert. Product will continue to ship so as to not impact critical patient care. BD recommends following Good Clinical Practice (GCP) by performing microscopy to confirm an elevated RBC Count in a urine sample as per CLSI GP16-3A.29(4) and CAP Checklist on Urinalysis.
- Quantity in commerce
- Total: 29,982,150 BD cups (US); 97,800,900 BD cups (OUS)
- Distribution
- Illinois
- Product codes and lots
- All lots within expiry
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0491-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0492-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0493-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0494-2019 | Class II | Terminated | Voluntary: Firm initiated |