Abbott i-STAT CG8+ Cartridges intended for use in the in vitro quantification of potassium in arterial...
Abbott Point Of Care Inc. began a recall of Abbott i-STAT CG8+ Cartridges intended for use in the in vitro quantification of potassium in arterial, venous, or capillary whole blood. List Number/UDI#... on . FDA classified it as Class II and gives the reason as "Inaccurate potassium results may be generated. When testing with control fluid and/or patient samples with low or reference range potassium concentration, elevated potassium results may be generated. When testing with...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0615-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Abbott i-STAT CG8+ Cartridges intended for use in the in vitro quantification of potassium in arterial, venous, or capillary whole blood. List Number/UDI# : 03P88-25 ,(01)10054749000163(17)181028(10)L18059
- Product code
- CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
- Reason for recall
- Inaccurate potassium results may be generated. When testing with control fluid and/or patient samples with low or reference range potassium concentration, elevated potassium results may be generated. When testing with samples with a high concentration of potassium, decreased potassium results may be generated
- Root cause tag
- manufacturing process
- FDA root cause
- Manufacturing material removal
- Action
- Abbott Point of Care issued recall letter on 9/13/18 via FedEx , advising of the problem, recommend action to return all unused cartridges to APOC . If you have any questions regarding this information, please contact Abbott Point of Care Technical Support at 844-256-9531 or your Abbott Point of Care representative.
- Quantity in commerce
- 17550 cartridges
- Distribution
- WI
- Product codes and lots
- Lot Number: L18059 Box Numbers: 0221, 0222, 0223, 0224, 0231, 0232, 0233
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0615-2019 | Class II | Terminated | Voluntary: Firm initiated |