Z-0107-2019Class II

Spectrum Iq Infusion System with Dose Iq Safety Software, 3570009

Baxter Healthcare Corporation / initiated 2018-09-18 / Terminated / root cause: software

Baxter Healthcare Corporation began a recall of SPECTRUM IQ INFUSION SYSTEM WITH DOSE IQ SAFETY SOFTWARE, 3570009 on . FDA classified it as Class II and gives the reason as "If an operator attempts to manually program a Basic Infusion for a drug that is not included in the Drug Library using the Basic-Dose Mode, the operator will receive a false hard limit exceeded error, preventing him or...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0107-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
Spectrum Iq Infusion System with Dose Iq Safety Software, 3570009
Product code
PHC Infusion Safety Management Software
Reason for recall
If an operator attempts to manually program a Basic Infusion for a drug that is not included in the Drug Library using the Basic-Dose Mode, the operator will receive a false hard limit exceeded error, preventing him or her from programming the infusion.
Root cause tag
software
FDA root cause
Software design
Action
The firm initiated their field correction by email on 09/14/2018. The notice alerted the consignee of the problem and stated that Baxter would be issuing a software update to correct the issue.
Quantity in commerce
354 units
Distribution
NJ
Product codes and lots
UDI 00085412610900 All Serial Numbers distributed prior to 07/09/2018

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0107-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0107-2019Class IITerminatedVoluntary: Firm initiated