Baxter Healthcare Corporation began a recall of SPECTRUM IQ INFUSION SYSTEM WITH DOSE IQ SAFETY SOFTWARE, 3570009 on . FDA classified it as Class II and gives the reason as "If an operator attempts to manually program a Basic Infusion for a drug that is not included in the Drug Library using the Basic-Dose Mode, the operator will receive a false hard limit exceeded error, preventing him or...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0107-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Spectrum Iq Infusion System with Dose Iq Safety Software, 3570009
- Product code
- PHC Infusion Safety Management Software
- Reason for recall
- If an operator attempts to manually program a Basic Infusion for a drug that is not included in the Drug Library using the Basic-Dose Mode, the operator will receive a false hard limit exceeded error, preventing him or her from programming the infusion.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm initiated their field correction by email on 09/14/2018. The notice alerted the consignee of the problem and stated that Baxter would be issuing a software update to correct the issue.
- Quantity in commerce
- 354 units
- Distribution
- NJ
- Product codes and lots
- UDI 00085412610900 All Serial Numbers distributed prior to 07/09/2018
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0107-2019 | Class II | Terminated | Voluntary: Firm initiated |