Z-0381-2019Class II
Liquid Cardiac Control, CQ5052
Randox Laboratories, Limited / initiated 2018-10-02 / Terminated
Randox Laboratories Ltd. began a recall of Liquid Cardiac Control, CQ5052 on . FDA classified it as Class II and gives the reason as "Randox has previously issued a recall for Liquid Cardiac Control CQ5053 lot 4245CK under REC334 on June 8, 2018. The firm has now confirmed the Liquid Cardiac Controls CQ5051, CQ5052 and CQ5053 are no longer suitable...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0381-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Randox Laboratories, Limited
- Product
- Liquid Cardiac Control, CQ5052
- Product code
- JJY Multi-Analyte Controls, All Kinds (Assayed)
- Reason for recall
- Randox has previously issued a recall for Liquid Cardiac Control CQ5053 lot 4245CK under REC334 on June 8, 2018. The firm has now confirmed the Liquid Cardiac Controls CQ5051, CQ5052 and CQ5053 are no longer suitable for the control of the Troponin T assay due to unacceptable variation between vials.
- FDA root cause
- Under Investigation by firm
- Action
- On October 2, 2018, the firm notified distributors of the recall via Expanded Urgent Medical Device Correction email. Customers were informed that the recall was related to previous recall REC334 on June 8, 2018. The firm has confirmed that Liquid Cardiac Controls CQ5052 and CQ5053 are no longer suitable for the control of the Troponin T assay due to unacceptable variation between vials. Customers were advised to take the following actions: - Discontinue use of these products for the Quality Control (QC) monitoring of Troponin T assays. - Discuss the contents of this notice with your Medical Director. - Update kits with revised IFUs excluding Troponin T values and the attached important notice to prevent further use of the device in the QC of Troponin T assays. - Complete and return the vigilance response section of this form to [email protected] within five working days. - Contact your local Randox sales representative for alternative product details. You may contact the firm at [email protected].
- Quantity in commerce
- 1738
- Distribution
- Distributed to IN, NY, PR, and WV.
- Product codes and lots
- All lots
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0380-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0381-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0382-2019 | Class II | Terminated | Voluntary: Firm initiated |