Z-0382-2019Class II

Liquid Cardiac Control, CQ5053

Randox Laboratories, Limited / initiated 2018-10-02 / Terminated

Randox Laboratories Ltd. began a recall of Liquid Cardiac Control, CQ5053 on . FDA classified it as Class II and gives the reason as "Randox has previously issued a recall for Liquid Cardiac Control CQ5053 lot 4245CK under REC334 on June 8, 2018. The firm has now confirmed the Liquid Cardiac Controls CQ5051, CQ5052 and CQ5053 are no longer suitable...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0382-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Randox Laboratories, Limited
Product
Liquid Cardiac Control, CQ5053
Product code
JJY Multi-Analyte Controls, All Kinds (Assayed)
Reason for recall
Randox has previously issued a recall for Liquid Cardiac Control CQ5053 lot 4245CK under REC334 on June 8, 2018. The firm has now confirmed the Liquid Cardiac Controls CQ5051, CQ5052 and CQ5053 are no longer suitable for the control of the Troponin T assay due to unacceptable variation between vials.
FDA root cause
Under Investigation by firm
Action
On October 2, 2018, the firm notified distributors of the recall via Expanded Urgent Medical Device Correction email. Customers were informed that the recall was related to previous recall REC334 on June 8, 2018. The firm has confirmed that Liquid Cardiac Controls CQ5052 and CQ5053 are no longer suitable for the control of the Troponin T assay due to unacceptable variation between vials. Customers were advised to take the following actions: - Discontinue use of these products for the Quality Control (QC) monitoring of Troponin T assays. - Discuss the contents of this notice with your Medical Director. - Update kits with revised IFUs excluding Troponin T values and the attached important notice to prevent further use of the device in the QC of Troponin T assays. - Complete and return the vigilance response section of this form to [email protected] within five working days. - Contact your local Randox sales representative for alternative product details. You may contact the firm at [email protected].
Quantity in commerce
2108
Distribution
Distributed to IN, NY, PR, and WV.
Product codes and lots
All lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0382-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0380-2019Class IITerminatedVoluntary: Firm initiated
Z-0381-2019Class IITerminatedVoluntary: Firm initiated
Z-0382-2019Class IITerminatedVoluntary: Firm initiated