Z-0377-2019Class II
Endoform Dermal Template 2x2, SKU 529311
AROA Biosurgery Inc. / initiated 2018-11-27 / Terminated / root cause: manufacturing process
AROA BIOSURGERY began a recall of Endoform Dermal Template 2x2, SKU 529311 Product Usage: Endoform Dermal Template is a sterile, single use ovine forestomach derived extracellular matrix... on . FDA classified it as Class II and gives the reason as "Potential for pouch seal failure". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0377-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- AROA Biosurgery Inc.
- Product
- Endoform Dermal Template 2x2, SKU 529311 Product Usage: Endoform Dermal Template is a sterile, single use ovine forestomach derived extracellular matrix intended to cover, protect and provide a moist wound environment.
- Product code
- KGN Wound Dressing With Animal-Derived Material(S)
- Reason for recall
- Potential for pouch seal failure
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- The firm sent a Customer Notification letter dated November 27, 2018. The letter identified the affected product, problem and actions to be taken. Customers are encouraged to visually inspect the seals on devices from these lots prior to use with particlar care. Should you observe a comprimised seal, please discard the device and select a new, fully sealed device for use.
- Quantity in commerce
- 8,853 total boxes
- Distribution
- US Nationwide Distribution
- Product codes and lots
- Lot numbers: EDT-7H06 EDT-7L01 EDT-7L03 EDT-8A01
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0377-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0378-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0379-2019 | Class II | Terminated | Voluntary: Firm initiated |