Z-0469-2019Class II

AIA-900 Analyzer, Product code 022930, 022930R

Tosoh Bioscience Inc / initiated 2018-10-05 / Terminated / root cause: design

Tosoh Bioscience Inc began a recall of AIA-900 Analyzer, Product code 022930, 022930R on . FDA classified it as Class II and gives the reason as "Tosoh Bioscience has become aware of a potential issue with the pickup arm and pickup arm motor on the AIA-900 Analyzer. If the pickup arm or pickup arm motor malfunctions, the analyzer will cease to function and result...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0469-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Tosoh Bioscience Inc
Product
AIA-900 Analyzer, Product code 022930, 022930R
Product code
KHO Fluorometer, For Clinical Use
Reason for recall
Tosoh Bioscience has become aware of a potential issue with the pickup arm and pickup arm motor on the AIA-900 Analyzer. If the pickup arm or pickup arm motor malfunctions, the analyzer will cease to function and result in an inability to test patient samples. This may cause a delay in testing and reporting patient results. Tosoh has received fifty-one (51) complaints related to this issue with no serious injuries reported.
Root cause tag
design
FDA root cause
Device Design
Action
On October 5, 2018, the firm notified its affected customers via an Urgent Medical Device Recall letter. The letter informed customers of the potential product issue with all active AIA-900 analyzers. The firm's service procedures have been updated to proactively troubleshoot and correct the pickup arm and pickup arm motor during installations and scheduled preventive maintenance visits. At the next scheduled preventive maintenance visit a Tosoh field service engineer will verify the pickup arm motor current and adjust as needed. Adherence to preventive maintenance schedules provides effective mitigation against potential failure of these components. Immediate Actions to be taken by the Customer/User " Thoroughly review the content of this letter. " Continue to use the instrument. A Tosoh representative will contact you to schedule your next preventive maintenance visit. o If a failure of the pickup arm is observed or the analyzer stops functioning, contact Tosoh Technical Support for an expedited service visit. " Complete and return the attached Acknowledgement Form by any of the listed methods. " File this notice with your laboratory records and forward this information to others in your laboratory. If you have obtained unexpected test results, delays in test result reporting, or received any complaints of illness or adverse events associated with the use of the analyzer, please contact Tosoh Technical Support 24 hours a day, seven days a week at (800) 248-6764. Should you have any questions regarding this medical device recall, please feel free to contact Bernadette O'Connell at (800) 248-6764 or by email at [email protected]. She will be available to answer any questions Monday - Friday, from 9:00 AM to 5:00 PM (PST).
Quantity in commerce
393
Distribution
Distributed nationwide. Foreign distribution to Canada and Latin America.
Product codes and lots
All units are affected

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0469-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0469-2019Class IITerminatedVoluntary: Firm initiated