Z-0446-2019Class II

-Outer Kit Box (P/N RFIT-LBL-0140) - FilmArray Meningitis/Encephalitis (ME) Panel REF RFIT-ASY-0118 30 For...

BioFire Diagnostics, LLC / initiated 2018-09-10 / Terminated / root cause: manufacturing process

BioFire Diagnostics, LLC began a recall of -Outer Kit Box (P/N RFIT-LBL-0140) - FilmArray Meningitis/Encephalitis (ME) Panel REF RFIT-ASY-0118 30 For FilmArray systems LOT KIT UDI (01)00815381020123 CE... on . FDA classified it as Class II and gives the reason as "Firm identified an issue in the manufacturing process which might contribute to elevated rates of false positive results for HSV-1 while using the ME Panel. Only specific kits are affected". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0446-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
BioFire Diagnostics, LLC
Product
-Outer Kit Box (P/N RFIT-LBL-0140) - FilmArray Meningitis/Encephalitis (ME) Panel REF RFIT-ASY-0118 30 For FilmArray systems LOT KIT UDI (01)00815381020123 CE 0086 IVD -Fitment Label (P/N RFIT-LBL-0138) which identifies the pouch label - FilmArray ME Panel v1.4 LOT SN CE 0086 IVD Product Usage: The FilmArray Meningitis/Encephalitis (ME) Panel is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with FilmArray systems.
Product code
PLO Meningitis/Encephalitis Pathogen Multiplex Nucleic Acid Detection System
Reason for recall
Firm identified an issue in the manufacturing process which might contribute to elevated rates of false positive results for HSV-1 while using the ME Panel. Only specific kits are affected.
Root cause tag
manufacturing process
FDA root cause
Material/Component Contamination
Action
On 09/10/2018, the firm sent an Urgent Medical Device Recall Letter via FedEx to the Recall Coordinator/Laboratory Director for all consignees. Consignees are instructed to: -Immediately examine their inventory for products with the affected Lot numbers and to discontinue use and to discard any remaining product in their possession that is subject to the recall.. BioFire will replace discarded product at no charge. -If the consignee has further distributed the affected product, they are to identify these customers and to notify them at once of the product recall. -Complete and return the Acknowledgement of Receipt Form that was enclosed with the Customer Advisory Notice letter, as soon as possible, and to return the Acknowledgement Form to the firm. Consignees were provided with firm contact information including phone numbers: 801-736-6354 and 1-800-735-6544
Quantity in commerce
1350 pouches
Distribution
Worldwide Distribution - US Nationwide in the states of: CA DC, FL, IN, KY, MN, MO, NB, NJ, OH, TX and WI. International: Canada, Columbia and Mexico
Product codes and lots
ME Panel Kit Lot No. ME Pouch Lot. No. Kit Expiration Date 580818 892018 4/2/2019 584418 896018 4/5/2019  Note - Kits contain a Primer Pill Lot 210318. The Primer Pill is a freeze dried/lyophilized "pellet" that contains all of the primers and buffer used in the PCR1 reaction.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0446-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0446-2019Class IITerminatedVoluntary: Firm initiated