Z-0483-2019Class II

Panda Freestanding Warmer. Device sold under the following product names: Panda Freestanding Warmer and Panda...

GE Healthcare / initiated 2018-09-25 / Terminated / root cause: manufacturing process

GE Healthcare, LLC began a recall of Panda Freestanding Warmer. Device sold under the following product names: Panda Freestanding Warmer and Panda Freestanding Warmer Generic ATO Model. Infant... on . FDA classified it as Class II and gives the reason as "Potential electric leakage affecting the user". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0483-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
Panda Freestanding Warmer. Device sold under the following product names: Panda Freestanding Warmer and Panda Freestanding Warmer Generic ATO Model. Infant Radiant warmer.
Product code
FMT Warmer, Infant Radiant
Reason for recall
Potential electric leakage affecting the user.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
GE Healthcare notified customers on about 09/25/2018 via URGENT MEDICAL DEVICE CORRECTION letter. The letter indicated that the device could continue to be used along with good clinical practices. GE Healthcare will visit each location to perform the necessary testing. Until GE Healthcare can visit each site, instructions are to disconnect devices that use the USB (Giraffe OmniBed Carestation and Giraffe Incubator Carestation only) and the RS232 Nurse call ports, and inspect all power cords to ensure there is no damage to the cord or ground pin. GE Healtchare will perform necessary electrical safety test on all affected products and correct if required. A GE Healthcare representative will contact each customer to arrange for this test. For questions or concerns, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Quantity in commerce
1,724 devices total
Distribution
Worldwide distribution. US Nationwide, Algeria, Australia, Austria, Belgium, Canada, Chile, China, Costa Rica, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Honduras, India, Indonesia, Ireland, Italy, Japan, Kazakhstan, Korea, Lithuania, Mexico, Netherlands, New Zealand, Norway, Pakistan, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, United Arab Emirates, United Kingdom, Vietnam.
Product codes and lots
Model Number, UDI (Serial Number): 2063568-001-535551, 010084068210391621PFSX7006711180626 (PFSX70067);  2063568-001-289325, Not Applicable (HDJT51400);  2063568-001-289325, Not Applicable (HDJT51401);  2063568-001-528373, 010084068210391621PFSX7005911180618 (PFSX70059);  2063568-001-529111, 010084068210391621PFSX7008111180731 (PFSX70081);  2063568-001-501555, 010084068210391621PFSV6011411161219 (PFSV60114);  2063568-001-562946, 010084068210391621PFSX7005011180503 (PFSX70050);  2063568-001-297793, Not Applicable (HDJT51542);  2063568-001-533799, 010084068210391621PFSW60069170515 (PFSW60069);  2063568-001-347865, Not Applicable (HDJT61189);  2063568-001-533888, 010084068210391621PFSW6007111170515 (PFSW60071);  2063568-001-533888, 010084068210391621PFSW6007011170515 (PFSW60070);  2063568-001-533888, 010084068210391621PFSW6007211170515 (PFSW60072);

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0483-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0479-2019Class IITerminatedVoluntary: Firm initiated
Z-0480-2019Class IITerminatedVoluntary: Firm initiated
Z-0481-2019Class IITerminatedVoluntary: Firm initiated
Z-0482-2019Class IITerminatedVoluntary: Firm initiated
Z-0483-2019Class IITerminatedVoluntary: Firm initiated
Z-0484-2019Class IITerminatedVoluntary: Firm initiated