Z-0468-2019Class II

Arrow AGB+ Multi-Lumen CVC Kit, 7 Fr., 3 Lumen, 20cm catheter length, .025 inch dia. spring-wire guide...

Arrow International LLC (subsidiary of Teleflex Incorporated) / initiated 2018-09-21 / Terminated / root cause: manufacturing process

Arrow International Inc began a recall of Arrow AGB+ Multi-Lumen CVC Kit, 7 Fr., 3 Lumen, 20cm catheter length, .025 inch dia. spring-wire guide, Product Code CDC-45703-B1A - Product Usage: The... on . FDA classified it as Class II and gives the reason as "The products labeling contains inconsistencies. In certain instances, the lidstock, banner card and shipper label state that the Spring Wire Guide (SWG) within is .032 rather than the correct SWG size .025 . For other...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0468-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Arrow International LLC (subsidiary of Teleflex Incorporated)
Product
Arrow AGB+ Multi-Lumen CVC Kit, 7 Fr., 3 Lumen, 20cm catheter length, .025 inch dia. spring-wire guide, Product Code CDC-45703-B1A - Product Usage: The multiple-lumen catheter permits venous access to the central circulation by the way of the femoral, jugular, or subclavian veins. The ARROWg+ard Blue Plus antimicrobial catheter is intended to help provide protection against catheter-related infections. It is not intended to be used as a treatment for existing infections nor is it indicated for long-term use.
Product code
FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
Reason for recall
The products labeling contains inconsistencies. In certain instances, the lidstock, banner card and shipper label state that the Spring Wire Guide (SWG) within is .032 rather than the correct SWG size .025 . For other lots in scope the lidstock correctly states the SWG size .025 however the banner card states the incorrect SWG size .032 .
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Urgent Medical Device Notification letters dated 9/21/18 were distributed to customers.
Quantity in commerce
3640
Distribution
US Nationwide distribution in the states of CA, FL, ID, IL, MI, NC, NJ, PA, and WI.
Product codes and lots
Lot/Batch Numbers: 13F17E0685 13F17F0562 13F17H0141 13F17K0340 13F18B0119 13R17E0685

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0468-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0468-2019Class IITerminatedVoluntary: Firm initiated