Z-0527-2019Class II

Optional MDRC-1119 Front Cover Attachment Kit for review display monitors

Barco N.V. / initiated 2017-03-30 / Terminated / root cause: design

Barco N.V. began a recall of Optional MDRC-1119 Front Cover Attachment Kit for review display monitors. Product Usage: The product is an optional kit, intended to be attached in front of... on . FDA classified it as Class II and gives the reason as "If metal strips are removed and re-applied to another display, there is potential for the metal strips to be deformed resulting in a smaller contact surface with the front cover magnet strips. There is a risk that the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0527-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Barco N.V.
Product
Optional MDRC-1119 Front Cover Attachment Kit for review display monitors. Product Usage: The product is an optional kit, intended to be attached in front of the LCD panel of a review display (MDRC-1119).
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
If metal strips are removed and re-applied to another display, there is potential for the metal strips to be deformed resulting in a smaller contact surface with the front cover magnet strips. There is a risk that the protective front cover could come loose and may fall down.
Root cause tag
design
FDA root cause
Device Design
Action
Customers were notified via email on about 03/30/2017. Customers were alerted that a safety related update has been made to the installation manual. A safety notice has been shipped to distributors with the updated installation instructions manual and an acknowledgement letter. Distributors were asked to do the following: a) Inform all your customers who have purchased the Front Cover Kit by sending them the Safety Notice; b) Send the updated installation instructions to all your customers who have purchased the Front Cover Kit; c) Fill out the acknowledgment letter; attach and send in reply to this email.
Quantity in commerce
2,087 units
Distribution
Worldwide - US Nationwide Distribution to CA, IL, GA, TX, and WI. In the countries of Belgium, Brazil, Canada, Czech Republic, Denmark, Finland, France, Germany, Italy, Japan, Poland, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Taiwan, Turkey.
Product codes and lots
Part Numbers C9826108, C9826109, C9826096, C9826097

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0527-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0527-2019Class IITerminatedVoluntary: Firm initiated