RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6 and RayStation 7, a stand-alone software treatment...
RAYSEARCH LABORATORIES AB began a recall of RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6 and RayStation 7, a stand-alone software treatment planning system on . FDA classified it as Class II and gives the reason as "Robust optimization for plans using beam set + background dose is not supported in RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6 and RayStation 7". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0294-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- RaySearch Laboratories AB (publ)
- Product
- RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6 and RayStation 7, a stand-alone software treatment planning system.
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- Robust optimization for plans using beam set + background dose is not supported in RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6 and RayStation 7.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Ray Search Labs notified userson 9/21/18 by email advising users of the action to be taken: After creating an adaptive plan from a plan with robust objectives, remove all robust objectives that are automatically added to the adaptive plan and replace them with other appropriate objectives before optimizing the adapted plan. o Do not use an optimization function template with robust functions for a dependent beam set.o Do not use scripting to add robust optimization functions related to beam set + background dose. Reply form to be completed/signed and returned. Regulatory information, please contact [email protected]
- Quantity in commerce
- 110 units
- Distribution
- Nationwide
- Product codes and lots
- Software, include version number: 4.5.0.19 , 4.5.1.14 , 4.5.2.7 , 4.7.0.15 , 4.7.1.10, 4.7.2.5 , 4.7.3.13 , 4.7.4.4 , 4.7.5.4 , 5.0.0.37 , 5.0.1.11 , 5.0.2.3 , 6.0.0.24 , 6.1.0.26 , 6.1.1.2 , 6.2.0.7 , 7.0.0.19
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0294-2019 | Class II | Terminated | Voluntary: Firm initiated |