Medtronic CareLink 2090 Programmer
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) began a recall of Medtronic CareLink 2090 Programmer on . FDA classified it as Class II and gives the reason as "Vulnerabilities have been identified in the SDN download process that may allow an individual with malicious intent to update the programmers with non-Medtronic software during an SDN download". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0406-2019
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic, Inc.
- Product
- Medtronic CareLink 2090 Programmer
- Product code
- OSR Pacemaker/Icd/Crt Non-Implanted Components
- Reason for recall
- Vulnerabilities have been identified in the SDN download process that may allow an individual with malicious intent to update the programmers with non-Medtronic software during an SDN download.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm began disseminating the Field Corrective Action notices by hand delivery on 10/02/2018 and followed with the posting of Medtronic Security Bulletin and the Urgent Medical Device Correction customer letter online on 10/11/2018. The notice explained vulnerabilities have been identified in the SDN download process that may allow an individual with malicious intent to update the programmers with non-Medtronic software during an SDN download and beginning October 11th, 2018, Medtronic will be disabling the SDN for programmer updates and will rely solely on the USB update method.
- Quantity in commerce
- 30172 units
- Distribution
- Worldwide
- Product codes and lots
- All Serial Numbers
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0406-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0407-2019 | Class II | Ongoing | Voluntary: Firm initiated |