Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0463-2019Class I

GE HEALTHCARE CARESCAPE R860 Inspiratory Safety Guard, Part Numbers: a) 2066713-001 (single pack) b)...

GE Healthcare / initiated 2018-10-05 / Terminated / root cause: manufacturing process

GE Healthcare, LLC began a recall of GE HEALTHCARE CARESCAPE R860 Inspiratory Safety Guard, Part Numbers: a) 2066713-001 (single pack) b) 2083208-001 (box of ten singles) The CARESCAPE R860... on . FDA classified it as Class I and gives the reason as "The ventilator Inspiratory Safety Guard (ISG) may disconnect from the breathing circuit pathway. As a result, this could create a loss of ventilation which may lead to inadequate oxygenation for patients, increasing the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0463-2019
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
GE HEALTHCARE CARESCAPE R860 Inspiratory Safety Guard, Part Numbers: a) 2066713-001 (single pack) b) 2083208-001 (box of ten singles) The CARESCAPE R860 ventilator is designed to provide mechanical ventilation or support to neonatal, pediatric, and adult patients weighing 0.25 kg and above. The CARESCAPE R860 ventilator is a microprocessor based, electronically controlled, pneumatically driven ventilator that includes integrated monitoring of FiO2, airway pressure, flow, and volume.
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
The ventilator Inspiratory Safety Guard (ISG) may disconnect from the breathing circuit pathway. As a result, this could create a loss of ventilation which may lead to inadequate oxygenation for patients, increasing the possibility of hypoxia.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The firm, GE Healthcare, initiated an "URGENT MEDICAL DEVICE CORRECTION" on 10/05/2018 by letter to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: Follow the safety instructions and If you have any incorrect ISGs, return to GE Healthcare or destroy on site, and complete and return the attached Customer Response form checking box #2 to indicate that you do have affected ISGs. E-mail the completed form to [email protected]. Please ensure that all potential users in your facility are made aware of this safety notification and the recommended actions. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative. The firm initiated a new recall notice identifying a new issue on 02/01/2019
Quantity in commerce
1140 units (307 units in US)
Distribution
Worldwide Distribution: US (Nationwide) including Puerto Rico, and to countries of: Albania, Argentina, Australia, Austria, Bangladesh, Bolivia, Bosnia and Herzegovina, Brazil, Cambodia, Canada, Chile, China, Colombia, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, Germany, Greece, Honduras, Hungary, India, Indonesia, Iraq, Italy, Japan, Kazakhstan, Korea (Republic of), Kuwait, Latvia, Lebanon, Libya, Malaysia, Maldives, Mexico, Myanmar, Nepal, Netherlands, Nicaragua, Nigeria, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, Uruguay, and Vietnam.
Product codes and lots
a) 2066713-001 Lot Numbers: 17/00951, 17/01174, 17/01937, 17/01994, 17/02372, 17/02393, 18/00126, 18/00127, 18/00128, 18/00129, 18/00130 b) 2083208-001 Lot Numbers: 17/00951, 17/01174, 17/01937, 17/01994, 17/02372, 17/02393, 18/00126, 18/00127, 18/00128, 18/00129, 18/00130

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0463-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0463-2019Class ITerminatedVoluntary: Firm initiated