Medtronic Navigation 0-arm 02 Imaging System interventional fluoroscopic x-ray system: OARM BI70002000 SYS...
Medtronic Navigation, Inc.-Littleton began a recall of Medtronic Navigation 0-arm 02 Imaging System interventional fluoroscopic x-ray system: OARM BI70002000 SYS IMAGING 02 Model: BI70002000 OARM BI70002000R SYS... on . FDA classified it as Class II and gives the reason as "O-arm 02 Imaging Systems correction to software version 4.1.0; new version of the user manual and an Image Acquisition System ("IAS") product label; replacing blown fuse of the new AC power input circuit". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0404-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Navigation, Inc
- Product
- Medtronic Navigation 0-arm 02 Imaging System interventional fluoroscopic x-ray system: OARM BI70002000 SYS IMAGING 02 Model: BI70002000 OARM BI70002000R SYS IMAGING 02 REFURB Model: BI70002000R
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- O-arm 02 Imaging Systems correction to software version 4.1.0; new version of the user manual and an Image Acquisition System ("IAS") product label; replacing blown fuse of the new AC power input circuit
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Medtronic Navigation issued Urgent Medical Device Correction letter on 9/27/18, notification letter advising, Medtronic field service personnel will install 4.1.0 software onto the designated consignee systems, update IFU and product labels as well as install a new designed AC power input circuit within the Mobile View Station {MVS). Questions regarding contact Medtronic Technical Support at 1-800-595-9709.
- Quantity in commerce
- 641 units
- Distribution
- Worldwide Distribution - US Nationwide Foreign: AUSTRALIA AUSTRIA BAHRAIN BANGLADESH BELGIUM BRAZIL CANADA CHILE COLOMBIA DENMARK FINLAND FRANCE GERMANY HR ICELAND INDIA IRELAND ISRAEL ITALY JAPAN JORDAN KOREA, REPUBLIC OF LUXEMBOURG MARTINIQUE MEXICO NETHERLANDS NEW ZEALAND POLAND PORTUGAL PUERTO RICO QATAR SAUDI ARABIA SINGAPORE SOUTH AFRICA SPAIN SWEDEN SWITZERLAND TAIWAN THAILAND TURKEY UNITED ARAB EMIRATES UNITED KINGDOM
- Product codes and lots
- Serial Numbers:C0875 through C2012
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0404-2019 | Class II | Terminated | Voluntary: Firm initiated |