Z-0404-2019Class II

Medtronic Navigation 0-arm 02 Imaging System interventional fluoroscopic x-ray system: OARM BI70002000 SYS...

Medtronic Navigation, Inc / initiated 2018-09-17 / Terminated / root cause: software

Medtronic Navigation, Inc.-Littleton began a recall of Medtronic Navigation 0-arm 02 Imaging System interventional fluoroscopic x-ray system: OARM BI70002000 SYS IMAGING 02 Model: BI70002000 OARM BI70002000R SYS... on . FDA classified it as Class II and gives the reason as "O-arm 02 Imaging Systems correction to software version 4.1.0; new version of the user manual and an Image Acquisition System ("IAS") product label; replacing blown fuse of the new AC power input circuit". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0404-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Navigation, Inc
Product
Medtronic Navigation 0-arm 02 Imaging System interventional fluoroscopic x-ray system: OARM BI70002000 SYS IMAGING 02 Model: BI70002000 OARM BI70002000R SYS IMAGING 02 REFURB Model: BI70002000R
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
O-arm 02 Imaging Systems correction to software version 4.1.0; new version of the user manual and an Image Acquisition System ("IAS") product label; replacing blown fuse of the new AC power input circuit
Root cause tag
software
FDA root cause
Software design
Action
Medtronic Navigation issued Urgent Medical Device Correction letter on 9/27/18, notification letter advising, Medtronic field service personnel will install 4.1.0 software onto the designated consignee systems, update IFU and product labels as well as install a new designed AC power input circuit within the Mobile View Station {MVS). Questions regarding contact Medtronic Technical Support at 1-800-595-9709.
Quantity in commerce
641 units
Distribution
Worldwide Distribution - US Nationwide Foreign: AUSTRALIA AUSTRIA BAHRAIN BANGLADESH BELGIUM BRAZIL CANADA CHILE COLOMBIA DENMARK FINLAND FRANCE GERMANY HR ICELAND INDIA IRELAND ISRAEL ITALY JAPAN JORDAN KOREA, REPUBLIC OF LUXEMBOURG MARTINIQUE MEXICO NETHERLANDS NEW ZEALAND POLAND PORTUGAL PUERTO RICO QATAR SAUDI ARABIA SINGAPORE SOUTH AFRICA SPAIN SWEDEN SWITZERLAND TAIWAN THAILAND TURKEY UNITED ARAB EMIRATES UNITED KINGDOM
Product codes and lots
Serial Numbers:C0875 through C2012

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0404-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0404-2019Class IITerminatedVoluntary: Firm initiated