Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0476-2019Class II

VITROS Chemistry Products dLDL Reagent-GENs 26 & 27. In Vitro Diagnostic to quantitatively measure LDL...

Ortho-Clinical Diagnostics, Inc. / initiated 2018-10-08 / Terminated

Ortho-Clinical Diagnostics began a recall of VITROS Chemistry Products dLDL Reagent-GENs 26 & 27. In Vitro Diagnostic to quantitatively measure LDL Cholesterol (LDLC) Product Code: 6801728 UDI... on . FDA classified it as Class II and gives the reason as "VITROS Chemistry Products dLDL Reagent GENs 26 and 27, negatively biased results". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0476-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS Chemistry Products dLDL Reagent-GENs 26 & 27. In Vitro Diagnostic to quantitatively measure LDL Cholesterol (LDLC) Product Code: 6801728 UDI (10758750006793)
Product code
MRR System, Test, Low Density, Lipoprotein
Reason for recall
VITROS Chemistry Products dLDL Reagent GENs 26 and 27, negatively biased results
FDA root cause
Under Investigation by firm
Action
Ortho Clinical Diagnostics issue the customer notification on October 8, 2018, provides information regarding the potential for biased results when using VITROS dLDL Reagent, Generations (GENs) 26 and 27. Actions to take: For GEN 26: Discontinue using GEN 26 and switch to an alternate GEN of product as soon as possible. Please specify full or partial sales units remaining and indicate credit or replacement on your Confirmation of Receipt form. " For GEN 27: Install ADD DRV 6003 on your VITROS System upon receipt. Following a successful recalibration, resume using VITROS dLDL Reagent, GEN 27 remaining in your inventory. " Complete the Confirmation of Receipt form and return by October 22, 2018. Questions contact our Ortho Care" Technical Solutions Center at 1-800-421-3311
Quantity in commerce
5850 units
Distribution
Nationwide Foreign: Australia Belgium Bermuda, HM11 Brazil Canada L3R 4G5 Chile China Colombia Denmark France Germany India Italy Japan Mexico Norway Poland Portugal Russia Singapore Spain Sweden The Netherlands United Kingdom
Product codes and lots
Lot Numbers:   1517-26-6396 (expires 6-Nov-2018)  1517-26-6582 (expires 2-Feb-2019)  1517-26-6639 (expires 2-Feb-2019)  1517-27-6808 (expires 7-May-2019)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0476-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0476-2019Class IITerminatedVoluntary: Firm initiated