AF w/AS 550 x 1-14ML/HR CPNB, Product code MVBXL- CPNB, Arrow AutoFuser Disposable Pain Control Pump
Arrow International Inc began a recall of AF w/AS 550 x 1-14ML/HR CPNB, Product code MVBXL- CPNB, Arrow AutoFuser Disposable Pain Control Pump on . FDA classified it as Class II and gives the reason as "Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0940-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Arrow International LLC (subsidiary of Teleflex Incorporated)
- Product
- AF w/AS 550 x 1-14ML/HR CPNB, Product code MVBXL- CPNB, Arrow AutoFuser Disposable Pain Control Pump
- Product code
- MEB Pump, Infusion, Elastomeric
- Reason for recall
- Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On October 2, 2018, the firm notified consignees via an Urgent Medical Device Recall letter. The letter informed customers of the product issue. Customers were instructed to take the following actions: 1. If you have affected stock in inventory, immediately discontinue use and quarantine the products. 2. Please complete the enclosed Recall Acknowledgement Form as soon as possible indicating whether you do or do not have stock and fax it to 1-855-419-8507 or email to [email protected]. A customer service representative will contact you with a Return Goods Authorization (RGA) number and provide instructions for the return of products. If you have any other questions, feel free to contact your local sales representative or Teleflex Customer Service at 1-866-396-2111.
- Quantity in commerce
- 38
- Distribution
- Nationwide distribution.
- Product codes and lots
- 74A1801376 74G1702799 74A1801900 74H1700167 74A1802544 74H1701111 74B1700925 74H1701796 74C1701765 74H1702406 74D1700392 74J1700193 74D1701253 74J1700680 74D1701539 74J1701583 74D1702241 74J1702180 74E1700485 74J1702742 74E1701026 74K1700635 74E1701692 74K1701435 74E1702410 74K1702040 74F1700147 74K1702604 74F1701966 74L1700203 74F1702681 74L1700932 74F1703240 74L1702112 74G1700539 74M1700833 74G1701226 74M1701615
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0940-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0941-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0942-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0943-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0944-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0945-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0946-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0947-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0948-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0949-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0950-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0951-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0952-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0953-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0954-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0955-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0956-2019 | Class II | Terminated | Voluntary: Firm initiated |