Sedecal NOVA FA DR System. for diagnostic radiography
Sedecal S.A. began a recall of Sedecal NOVA FA DR System. for diagnostic radiography on . FDA classified it as Class II and gives the reason as "A safety problem with installation and maintenance could lead to the breakage of the two steel cables that support the equipment to its roof anchor. Due to this, the equipment may fall and cause harm to the patient...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0488-2019
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Sedecal S.A.
- Product
- Sedecal NOVA FA DR System. for diagnostic radiography.
- Product code
- MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)
- Reason for recall
- A safety problem with installation and maintenance could lead to the breakage of the two steel cables that support the equipment to its roof anchor. Due to this, the equipment may fall and cause harm to the patient, user, or third parties.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Initial communication from SEDECAL to all our distributors was performed on 07/18/2018 by email. The letter explained the problem and requested the consignee cease use of the unit if they observe an abnormal displacement in the vertical axis or any suspicious noise during the vertical movement.
- Quantity in commerce
- 145 units
- Distribution
- US and Canada
- Product codes and lots
- Serial Numbers 07360816 through 17291024
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0488-2019 | Class II | Ongoing | Voluntary: Firm initiated |