Z-0488-2019Class II

Sedecal NOVA FA DR System. for diagnostic radiography

Sedecal S.A. / initiated 2018-07-18 / Open, Classified / root cause: design

Sedecal S.A. began a recall of Sedecal NOVA FA DR System. for diagnostic radiography on . FDA classified it as Class II and gives the reason as "A safety problem with installation and maintenance could lead to the breakage of the two steel cables that support the equipment to its roof anchor. Due to this, the equipment may fall and cause harm to the patient...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0488-2019
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Sedecal S.A.
Product
Sedecal NOVA FA DR System. for diagnostic radiography.
Product code
MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)
Reason for recall
A safety problem with installation and maintenance could lead to the breakage of the two steel cables that support the equipment to its roof anchor. Due to this, the equipment may fall and cause harm to the patient, user, or third parties.
Root cause tag
design
FDA root cause
Device Design
Action
Initial communication from SEDECAL to all our distributors was performed on 07/18/2018 by email. The letter explained the problem and requested the consignee cease use of the unit if they observe an abnormal displacement in the vertical axis or any suspicious noise during the vertical movement.
Quantity in commerce
145 units
Distribution
US and Canada
Product codes and lots
Serial Numbers 07360816 through 17291024

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0488-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0488-2019Class IIOngoingVoluntary: Firm initiated