Z-0799-2019Class II

ARCHITECT Total T3 Reagent Kit used on ARCHITECT i1000SR or i2000/i2000SR platforms

Abbott Ireland Diagnostics Division / initiated 2018-10-12 / Terminated / root cause: design

Abbott Ireland Diagnostics Division began a recall of ARCHITECT Total T3 Reagent Kit used on ARCHITECT i1000SR or i2000/i2000SR platforms Product Usage: It is a Chemiluminescent Microparticle Immunoassay (CMIA)... on . FDA classified it as Class II and gives the reason as "Samples tested using ARCHITECT Free T3 or ARCHITECT Total T3 assays may show depressed results due to a reagent carryover when testing on board with specific assays on the ARCHITECT i1000SR and i2000/i2999SR platforms". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0799-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Ireland Diagnostics Division
Product
ARCHITECT Total T3 Reagent Kit used on ARCHITECT i1000SR or i2000/i2000SR platforms Product Usage: It is a Chemiluminescent Microparticle Immunoassay (CMIA) for the quantitative determination of free triiodothyronine (Free T3) in human serum and plasma.
Product code
CDP Radioimmunoassay, Total Triiodothyronine
Reason for recall
Samples tested using ARCHITECT Free T3 or ARCHITECT Total T3 assays may show depressed results due to a reagent carryover when testing on board with specific assays on the ARCHITECT i1000SR and i2000/i2999SR platforms.
Root cause tag
design
FDA root cause
Device Design
Action
Abbott notified customers on about 10/12/218 via "Product Correction Urgent - Immediate Action Required" letter. Letters were sent separately to customers who utilize the ARCHITECH i1000SR platform and the i2000/i2000SR platforms. Necessary actions for customers who utilize the ARCHITECH i1000SR platform included that if a separate instrument can be used, then separate the ARCHITECT Free T3 and ARCHITECT Total T3 assays from the specific assays listed in the letter (ARCHITECT TSH, ARCHITECT TUptake, ARCHITECT HIV Ag/Ab Combo, ARCHITECT Cortisol, ARCHITECT LH, ARCHITECT ProGRP, ARCHITECT rHTLVI/II, ARCHITECT Total PSA, ARCHITECT AFP, ARCHITECT Free PSA, ARCHITECT 25OH Vitamin D), by running these tests on different instruments. For those customers who cannot use a separate instrument, then they are instructed to perform the Maintenance Procedure (6445 Pipettor/WZ Probe Cleaning for i1000SR) in the i1000SR Operations on the instrument prior to performing batch testing for all ARCHITECT Free T3 or ARCHITECT Total T3 samples. Necessary actions for customers who utilize the i2000/i2000SR platforms included that if a separate instrument can be used, then separate the ARCHITECT Free T3 and ARCHITECT Total T3 assays from the ARCHITECT 25OH Vitamin D assay, by running these tests on different instruments. For those customers who cannot use a separate instrument, then perform Daily Maintenance (Operations Manual, 6041 for i2000/i2000SR) on the instrument prior to performing batch testing for all ARCHITECT Free T3 or ARCHITECT Total T3 samples. Additionally, instructions for both types of affected platforms included to review the letter with each customer site's Medical Director or Laboratory Director and follow laboratory protocol regarding the need for reviewing previously reported patient results. If the product has been forwarded to other laboratories, please inform them of the recall and provide a copy of the notification letter. Also, complete and r
Quantity in commerce
191,410 units total
Distribution
Worldwide Distribution - US Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NJ, NV, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, PR. International distribution to AFGHANISTAN, ALBANIA, ALGERIA, ANGOLA, ARMENIA, AUSTRIA, AZERBAIJAN, BAHAMAS, BAHRAIN, BANGLADESH, BELARUS, BELGIUM, BOSNIA AND HERZEGOVI, BRUNEI, BULGARIA, BURKINA FASO, CAMEROON, CANADA, CAYMAN ISLANDS, CHILE, CROATIA, CYPRUS, CZECH REPUBLIC, DEM. REP. OF THE CON, DENMARK, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, ETHIOPIA, FINLAND, FRANCE, GAZA & JERICHO, GEORGIA, GERMANY, GHANA, GREECE, HONDURAS, HUNGARY, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, IVORY COAST, JAMAICA, JORDAN, KAZAKHSTAN, KENYA, KUWAIT, KYRGYZSTAN, LATVIA, LEBANON, LIBYA, LITHUANIA, LUXEMBOURG, MACEDONIA, MALAWI, MALAYSIA, MALDIVES, MALI, MAURITANIA, MAURITIUS, MOLDOVA, MONTENEGRO, MOROCCO, MOZAMBIQUE, NEPAL, NETHERLANDS, NICARAGUA, NIGERIA, NORWAY, OMAN, PAKISTAN, PANAMA, PARAGUAY, PERU, POLAND, PORTUGAL, QATAR, REPUBLIC OF YEMEN, ROMANIA, RUSSIA, RWANDA, SAUDI ARABIA, SENEGAL, SERBIA, SINGAPORE, SLOVAKIA,SLOVENIA, SOUTH AFRICA, SPAIN, SRI LANKA, ST. LUCIA, ST. VINCENT, SWEDEN, SWITZERLAND, TANZANIA, THE GAMBIA,TRINIDAD & TOBAGO, TUNISIA, TURKEY, UAE, UGANDA, UKRAINE,UNITED KINGDOM, VIETNAM, ZIMBABWE.
Product codes and lots
List number (Kit Configuration): 7K64 20 (4 x 100 test/kit), 7K64 25 (1 x 100 tests/kit), 7K64 30 (4 x 500 tests/kit), 7K64 35 (1 x 500 tests/kit), 7K64 27 (1 x 100 tests/kit), 7K64 32 (4 x 500 tests/kit), 7K64 37 (1 x 500 tests/kit). All lot codes for each of these list numbers are included in the recall.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0799-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0798-2019Class IITerminatedVoluntary: Firm initiated
Z-0799-2019Class IITerminatedVoluntary: Firm initiated