NEOCONNECT, Enteral Extension with ENFit connector, REF EXT-60NC The NeoMed NeoConnect Enteral Only Extension...
Neomed Inc began a recall of NEOCONNECT, Enteral Extension with ENFit connector, REF EXT-60NC The NeoMed NeoConnect Enteral Only Extension Set is intended for use as an extension set for... on . FDA classified it as Class III and gives the reason as "The sterile pouch label states the expiration date is , but the correct expiration date is . The case label and the shipper label have the correct date". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0526-2019
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Halyard Health
- Product
- NEOCONNECT, Enteral Extension with ENFit connector, REF EXT-60NC The NeoMed NeoConnect Enteral Only Extension Set is intended for use as an extension set for nasogastric/oralgastric or gastric tube enteral feeding tubes, incorporating safety connectors which help mitigate the risk of accidental connection of an I.V. system to the enteral system or the enteral system to an I.V. system. This device is indicated for use in neonatal and pediatric patients in connection with an enteral feeding tube to provide nutrition via nasal or oral gastric placements.
- Product code
- KNT Tubes, Gastrointestinal (And Accessories)
- Reason for recall
- The sterile pouch label states the expiration date is 2016-11-19, but the correct expiration date is 2019-11-09. The case label and the shipper label have the correct date
- Root cause tag
- labeling and UDI
- FDA root cause
- Under Investigation by firm
- Action
- The firm, NeoMed, sent an URGENT: MEDICAL DEVICE RECALL" letter dated 10/24/2018 to its consignees. The letter described the product, problem and actions to be taken. The consignees were instructed to do the following: Please contact NeoMed Quality Assurance by email to [email protected] to confirm receipt of this letter and obtain a Return Materials Authorization for any units identified above that remain in your inventory. Or, you may mail the enclosed reply card to NeoMed Inc., 100 Londonderry Ct #112, Woodstock, GA 30188. If you have any questions, call 770-516-2225 or email: [email protected].
- Quantity in commerce
- 654 boxes (32700 units)
- Distribution
- US Distribution to states of: California, Florida, Illinois, Michigan, New Jersey, Ohio, Tennessee, and Texas.
- Product codes and lots
- (UDIs) on the shipper [(01)20817584012015 (17)191109 (10)20161110], case [(01)10817584012018 (17)1991109 (10)20161110] and pouch [(01)00817584012011 (17)191109 (10)20161110] Lot Number: 20161110
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0526-2019 | Class III | Terminated | Voluntary: Firm initiated |