Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0474-2019Class II

The SQ-RX Pulse Generator a component of the Subcutaneous Implantable Cardioverter Defibrillator (S-ICD)...

Boston Scientific Corporation / initiated 2018-11-01 / Terminated / root cause: design

Boston Scientific Corporation began a recall of The SQ-RX Pulse Generator a component of the Subcutaneous Implantable Cardioverter Defibrillator (S-ICD) System The S-ICD System is intended to provide... on . FDA classified it as Class II and gives the reason as "Experienced accelerated battery depletion and a shortened replacement interval due to latent internal battery malfunctions". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0474-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
The SQ-RX Pulse Generator a component of the Subcutaneous Implantable Cardioverter Defibrillator (S-ICD) System The S-ICD System is intended to provide defibrillation therapy for the treatment of life-threatening ventricular tachyarrhythmias in patients who do not have symptomatic bradycardia, incessant ventricular tachycardia, or spontaneous, frequently recurring ventricular tachycardia that is reliably terminated with anti-tachycardia pacing.
Product code
LWS Implantable Cardioverter Defibrillator (Non-Crt)
Reason for recall
Experienced accelerated battery depletion and a shortened replacement interval due to latent internal battery malfunctions.
Root cause tag
design
FDA root cause
Device Design
Action
The firm, Boston Scientific, sent an "Important Medical Device Information" letter to customers on November 5, 2018 by hand, email via our sales representatives or Federal Express or Courier. The letter described the product, problem and actions to be taken. We will reconcile that 100% of the identified customers /customer representatives signed for the letter, or the letter has been confirmed to be delivered. Boston Scientific will post the product advisory letter on our website, within the Product Performance Resource Center located @ www.bostonscientific.com/en-US/pprc/product-advisories.html. The customers were instructed to do the following: Your doctor has been notified to look for this possible change, so please go to all scheduled defibrillator checks and clinic visits. As always, if you or your family hear beeping from your device, please contact your heart clinic as soon as possible. Your doctor can confirm your ability to hear and recognize this sound during an appointment. If you have additional questions regarding this information or would like to report a clinical event, please contact your Boston Scientific representative or Technical Services, or the Recall Contact at 651-582-3494 or email: [email protected].
Quantity in commerce
12,900
Distribution
US Nationwide Distribution: Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West Virginia, Wisconsin, and Wyoming.
Product codes and lots
SQ-RX Pulse Generator Model 1010

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0474-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0474-2019Class IITerminatedVoluntary: Firm initiated