HemoTrol Level 2- In Vitro Diagnostic Hematology quality control mixture Catalogue number: 171.002.002...
EUROTROL INC began a recall of HemoTrol Level 2- In Vitro Diagnostic Hematology quality control mixture Catalogue number: 171.002.002 Eurotrol HemoLin is an assayed hemoglobin control... on . FDA classified it as Class II and gives the reason as "Incorrect measurement results caused by microbial contamination of the product, results in values lower than assigned values". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0514-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Eurotrol B.V.
- Product
- HemoTrol Level 2- In Vitro Diagnostic Hematology quality control mixture Catalogue number: 171.002.002 Eurotrol HemoLin is an assayed hemoglobin control intended for professional use in the verification of the linearity of the HemoCue Hb 201 systems
- Product code
- GGM Control, Hemoglobin
- Reason for recall
- Incorrect measurement results caused by microbial contamination of the product, results in values lower than assigned values
- Root cause tag
- manufacturing process
- FDA root cause
- Material/Component Contamination
- Action
- Eurotrol distributors and initial consignees contacted via e-mail with the recall notification letter on 10/16/18. The recall notification letter sent via overnight mail in case e-mail is not deliverable. The recall notification letter instructs distributors how to perform the subrecall towards end-users and inform Eurotrol about the performed subrecall actions Primary instruction will be to stop using the product (discard according local regulation on medical waste) and contact Eurotrol for replacement material. Questions, contact (Monday through Friday, 9:00 AM to 5:00 PM, Eastern Time): Name: Eurotrol Inc. Address 850 N Black Branch Rd. Zipcode & city: Elizabethtown, KY 42701 USA Phone: T +1 502 501 1180 E-mail: [email protected].
- Quantity in commerce
- 2000
- Distribution
- Worldwide Distribution: US (Nationwide) and countries of: Austria, Australia, Bahrain, France, Hong Kong, Indonesia, Ireland, Japan, Jordan, Kenya, Republic of Korea, Kuwait, Maldives, New Zealand, Mozambique, Pakistan, Poland, Slovenia, Africa, Spain, Thailand, Finland, Norway, and UK.
- Product codes and lots
- Lot Number: 81166
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0514-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0515-2019 | Class II | Terminated | Voluntary: Firm initiated |