Biomet 20mm Compress Device Short Anchor Plug With Drill Set / With Pin Inserter Ion Implant Item number...
Zimmer Biomet, Inc. began a recall of Biomet 20mm Compress Device Short Anchor Plug With Drill Set / With Pin Inserter Ion Implant Item number: 178562 Product Usage: 1. Correction of revision of... on . FDA classified it as Class II and gives the reason as "Potential metal burrs in the holes of the device preventing the appropriate drill and transverse pins from being inserted". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0553-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Biomet 20mm Compress Device Short Anchor Plug With Drill Set / With Pin Inserter Ion Implant Item number: 178562 Product Usage: 1. Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. 2. Tumor resections. 3. Revision of previously failed total joint arthroplasty. 4. Trauma. The Compress Segmental Femoral Replacement System (Short Spindle and Anchor Plug) components are intended for uncemented use only.
- Product code
- JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented
- Reason for recall
- Potential metal burrs in the holes of the device preventing the appropriate drill and transverse pins from being inserted
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Zimmer Biomet initiated recall by letter on 10/24/18 to distributors and hospitals advising of the problem and action to be taken-locate quarantine and return product. Complete Inventory Return Certification Form. Questions or concerns call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday.
- Quantity in commerce
- 6 units
- Distribution
- US Nationwide Distribution in the states of IN, DC, MD
- Product codes and lots
- Lot Number: 608740 UDI Number: (01)00880304461666(17)271113(10)608740
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0553-2019 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY