Z-0457-2019Class II

ARTIS zee biplane systems with Material 10094141- a Angiographic x-ray system

Siemens Medical Solutions USA, Inc. / initiated 2018-10-19 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of ARTIS zee biplane systems with Material 10094141- a Angiographic x-ray system Product Usage: Artis zee is a dedicated angiography system developed for single... on . FDA classified it as Class II and gives the reason as "Vibrations of the C-arm of the plane B may result in reduced image quality caused by the screw connection to the C-arm in the ceiling support". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0457-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
ARTIS zee biplane systems with Material 10094141- a Angiographic x-ray system Product Usage: Artis zee is a dedicated angiography system developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis zee includes cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. Artis zee can also support the acquisition of position triggered imaging for spatial data synthesis.
Product code
IZI System, X-Ray, Angiographic
Reason for recall
Vibrations of the C-arm of the plane B may result in reduced image quality caused by the screw connection to the C-arm in the ceiling support
Root cause tag
design
FDA root cause
Device Design
Action
Siemens issued on 10/19/18 an Urgent Medial Devoce Correction notification to via AX033/18/S advising users of the problem, health risk, steps to be taken. Siemens has initiated screw connection exchange of the ceiling bearing via Update Instruction AXOSS/17 /S.
Quantity in commerce
14 units
Distribution
US Nationwide Distribution
Product codes and lots
Serial Numbers: Serial 155122 155149 155151 155105 155148 154671 155126 155131 155132 155119 155121 155118 155147

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0457-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0457-2019Class IITerminatedVoluntary: Firm initiated
Z-0458-2019Class IITerminatedVoluntary: Firm initiated
Z-0459-2019Class IITerminatedVoluntary: Firm initiated