ARTIS Q biplane with Material 10848282- - a Angiographic x-ray system
Siemens Medical Solutions USA, Inc began a recall of ARTIS Q biplane with Material 10848282- - a Angiographic x-ray system Product Usage: Artis zee is a dedicated angiography system developed for single and... on . FDA classified it as Class II and gives the reason as "Vibrations of the C-arm of the plane B may result in reduced image quality caused by the screw connection to the C-arm in the ceiling support". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0458-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- ARTIS Q biplane with Material 10848282- - a Angiographic x-ray system Product Usage: Artis zee is a dedicated angiography system developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis zee includes cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. Artis zee can also support the acquisition of position triggered imaging for spatial data synthesis.
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- Vibrations of the C-arm of the plane B may result in reduced image quality caused by the screw connection to the C-arm in the ceiling support
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Siemens issued on 10/19/18 an Urgent Medial Devoce Correction notification to via AX033/18/S advising users of the problem, health risk, steps to be taken. Siemens has initiated screw connection exchange of the ceiling bearing via Update Instruction AXOSS/17 /S.
- Quantity in commerce
- 54 units
- Distribution
- US Nationwide Distribution
- Product codes and lots
- Serial Numbers: 121336 121368 121208 121371 121374 121209 121325 121357 121378 121369 121364 121314 121351 121354 121340 121365 121324 121216 121333 121381 121352 121355 121210 121215 121379 121366 121323 121370 121384 121332 121344 121309 121305 121342 121301 121315 121372 121321 121383 121312 121359 121322 121356 121319 121211 121375 121377 121361 121339 121310 121362 121316 121212 121313
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0457-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0458-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0459-2019 | Class II | Terminated | Voluntary: Firm initiated |