ARTIS Q.zen biplane with Material 10848355 - a Angiographic x-ray system
Siemens Medical Solutions USA, Inc began a recall of ARTIS Q.zen biplane with Material 10848355 - a Angiographic x-ray system Product Usage: Artis zee is a dedicated angiography system developed for single and... on . FDA classified it as Class II and gives the reason as "Vibrations of the C-arm of the plane B may result in reduced image quality caused by the screw connection to the C-arm in the ceiling support". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0459-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- ARTIS Q.zen biplane with Material 10848355 - a Angiographic x-ray system Product Usage: Artis zee is a dedicated angiography system developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis zee includes cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. Artis zee can also support the acquisition of position triggered imaging for spatial data synthesis.
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- Vibrations of the C-arm of the plane B may result in reduced image quality caused by the screw connection to the C-arm in the ceiling support
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Siemens issued on 10/19/18 an Urgent Medial Devoce Correction notification to via AX033/18/S advising users of the problem, health risk, steps to be taken. Siemens has initiated screw connection exchange of the ceiling bearing via Update Instruction AXOSS/17 /S.
- Quantity in commerce
- 15
- Distribution
- US Nationwide Distribution
- Product codes and lots
- Serial Numbers: Serial 123202 123211 123208 123209 123206 123207 123219 123205 123201 123055 123056 123216 123200 123212
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0457-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0458-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0459-2019 | Class II | Terminated | Voluntary: Firm initiated |