Laser Fiber Assy-Single Use-LFS-272-RT-0.22NA-SMA905- Slim Extension Sleeve Silver-Sterile ETO- Private...
Leoni Fiber Optics GmbH began a recall of Laser Fiber Assy-Single Use-LFS-272-RT-0.22NA-SMA905- Slim Extension Sleeve Silver-Sterile ETO- Private label: SU-200-RT SmartScope Packaging. The Laser... on . FDA classified it as Class II and gives the reason as "A percentage of the products shipped in Smartscope single-box packaging may exhibit a broken package seal, thus having compromised sterility. There have been two reports of compromised sterility; however, all products...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0577-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Leoni Fiber Optics, Inc.
- Product
- Laser Fiber Assy-Single Use-LFS-272-RT-0.22NA-SMA905- Slim Extension Sleeve Silver-Sterile ETO- Private label: SU-200-RT SmartScope Packaging. The Laser peripherals bare laser fibers, ENT fibers and Endoprobes are intended for use in laser surgical procedure for cutting, coagulating, or vaporizing in any soft tissue application for which compatible Nd: YAG, Ho: YAG, KTP and Diode lasers have been cleared.
- Product code
- GEX Powered Laser Surgical Instrument
- Reason for recall
- A percentage of the products shipped in Smartscope single-box packaging may exhibit a broken package seal, thus having compromised sterility. There have been two reports of compromised sterility; however, all products packaged in Smartscope single-box packaging are at risk of a compromised product seal. Broken seal pouches compromise sterility and could increase the risk of infection if used.
- Root cause tag
- packaging
- FDA root cause
- Packaging
- Action
- The firm, LEONI, notified its distributor of the product issue on 10/16/18 via phone call and a subsequent "URGENT: MEDICAL DEVICE RECALL" letter. The distributor was informed of the product issue, and asked to disseminate recall information to customers. Actions to be taken by the Customer/User: End users should discontinue usage of the product. Product should be immediately quarantined. The Distributor should complete the acknowledgement of receipt form, to arrange for replacement and or credit to their account. A Leoni Customer Service Representative shall contact you with a Return material Authorization. If a replacement order is needed please inform customer service at the time of RMA issuance. - Discontinue usage if packaging irregularities exist - Return RMA units to Leoni Fiber Optics Inc. - Inform Customer Service if replacement product is required. - Complete acknowledge form and return back to Leoni Fiber Optics Inc. Attention Roger West, LFOI Quality Manager, [email protected] If you have any questions, please contact Quality Manager at 757-258-4805 or Email: [email protected].
- Distribution
- US Distribution to the state of: Ohio.
- Product codes and lots
- 096874 00816901020265
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0577-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0578-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0579-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0580-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0581-2019 | Class II | Terminated | Voluntary: Firm initiated |