Z-0592-2019Class II

IntelliVue MX40 Patient Monitor, Model Nos. 865350, 865351, 865352, 867146 -

Philips North America LLC / initiated 2018-10-10 / Open, Classified / root cause: supplier

Philips Electronics North America Corporation began a recall of IntelliVue MX40 Patient Monitor, Model Nos. 865350, 865351, 865352, 867146 - Product Usage: Indicated for use by health care professionals whenever there is a... on . FDA classified it as Class II and gives the reason as "A problem has been detected in the Philips IntelliVue MX4O that, if it were to occur, could affect the performance of the equipment. The volume of the MX4O speaker in these devices may be diminished or not audible above...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as supplier.

Recall record

Recall number
Z-0592-2019
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Philips North America LLC
Product
IntelliVue MX40 Patient Monitor, Model Nos. 865350, 865351, 865352, 867146 - Product Usage: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults and pediatrics in hospital environments and during transport inside hospitals.
Product code
DSI Detector And Alarm, Arrhythmia
Reason for recall
A problem has been detected in the Philips IntelliVue MX4O that, if it were to occur, could affect the performance of the equipment. The volume of the MX4O speaker in these devices may be diminished or not audible above 4,500 feet when operating in Monitor Mode. These devices may intermittently exhibit Speaker Malfunct INOP messages after the Power On Self-Test. The issue is only apparent when the device is being used in Monitor Mode at an altitude of greater than 4,500 feet.
Root cause tag
supplier
FDA root cause
Vendor change control
Action
On November 9, 2018, the firm notified customers of the issue via Customer Information Letters. The letter discussed the product problem. Customers are asked to confirm speaker behavior. It is recommended that the clinical user reboot the MX40 prior to entering Monitoring Mode to confirm proper operation of the speaker. Customers are also asked to complete the attached Reply Card and return to Philips as soon as possible. The firm will replace affected devices used at above 4,500 feet (1,372 meters). If you need any further information or support concerning this issue, please contact Philips Customer Support at 1-800-722-9377.
Quantity in commerce
35469 units
Distribution
US Nationwide distribution.
Product codes and lots
All devices manufactured or repaired from 21-FEB-2017 to 22-AUG -2018. (Serial numbers included as an attachment; too many to list.)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0592-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0592-2019Class IIOngoingVoluntary: Firm initiated