Salem Sump Dual Lumen Stomach Tube 10 Fr/Ch (3.3 mm) x 36" (91 cm), Product Code 8888264911
COVIDIEN LLC began a recall of Salem Sump Dual Lumen Stomach Tube 10 Fr/Ch (3.3 mm) x 36" (91 cm), Product Code 8888264911 Product Usage: The Salem Sump Tubes are intended for decompression... on . FDA classified it as Class II and gives the reason as "This voluntary recall is being conducted due to incorrect packaging. In one lot of Salem Sump Dual Lumen Stomach Tubes the 10Fr product was packaged in a 16Fr pouch. The use of a 10 Fr Salem Sump Tube when a 16Fr is...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0597-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Salem Sump Dual Lumen Stomach Tube 10 Fr/Ch (3.3 mm) x 36" (91 cm), Product Code 8888264911 Product Usage: The Salem Sump Tubes are intended for decompression and drainage of stomach contents following injury, illness, or surgery in patients with intestinal obstruction, and to prevent the distention of the stomach.
- Product code
- KNT Tubes, Gastrointestinal (And Accessories)
- Reason for recall
- This voluntary recall is being conducted due to incorrect packaging. In one lot of Salem Sump Dual Lumen Stomach Tubes the 10Fr product was packaged in a 16Fr pouch. The use of a 10 Fr Salem Sump Tube when a 16Fr is needed could result in decreased fluid collection, with the potential to affect patient care, particularly in an emergency situation where the correct size is not readily available.
- Root cause tag
- packaging
- FDA root cause
- Under Investigation by firm
- Action
- On October 31, 2018, the firm distributed Urgent Medical Device Recall letters to affected customers. Customers were informed of the incorrect packaging issue. Action Required: 1. INSPECT your inventory for the affected product code and lot number (listed on package) 2. SEGREGATE and QUARANTINE all on-hand product 3. RETURN the enclosed acknowledgment form via fax to 847-689-9101 or 614-652-9648, whether or not you have affected product, indicating the product code, lot and quantity of product youve quarantined or discarded. 4. NOTIFY any customers to whom you may have distributed, or forwarded product affected by this recall 5. CONTACT the appropriate Customer Service group to arrange for return and credit of any affected product: Hospital800-964-5227 Federal Government800-444-1166 Distributor800-635-6021 All other customers888-444-54401
- Quantity in commerce
- 1522
- Distribution
- Worldwide Distribution - US Nationwide and foreign distribution to Canada.
- Product codes and lots
- 1812115564