VITROS 5,1 FS Chemistry System, Version 3.0 & below For use in the in vitro quantitative measurement of a...

Ortho-Clinical Diagnostics, Inc. / initiated 2018-10-19 / Terminated / root cause: software

Ortho-Clinical Diagnostics began a recall of VITROS 5,1 FS Chemistry System, Version 3.0 & below For use in the in vitro quantitative measurement of a variety of analytes of clinical interest in certain... on . FDA classified it as Class II and gives the reason as "Software anomaly may incorrectly extend the on-analyzer stability time for a loaded VITROS Reagent". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0651-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS 5,1 FS Chemistry System, Version 3.0 & below For use in the in vitro quantitative measurement of a variety of analytes of clinical interest in certain biological fluids (depending on the analyte) such as serum, plasma, urine and cerebrospinal fluid.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Software anomaly may incorrectly extend the on-analyzer stability time for a loaded VITROS Reagent.
Root cause tag
software
FDA root cause
Software Manufacturing/Software Deployment
Action
On October 19, 2018, Ortho Clinical Diagnostics issued "URGENT PRODUCT CORRECTION NOTIFICATION" notices via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) to all customers to inform them of the issue and actions to take to prevent occurrence of the anomaly. On November 19, 2018 Ortho Clinical Diagnostics re-issued correction notices to customers pertaining to the 5,1 FS system (CL2018-194_5,1FS). Customers may continue using their VITROS Systems and have been informed that the software issue will be mitigated with the next version of software. The expected dates of these software versions are included in the communication. Foreign affiliates were informed by email approximately on 21 October 2018 of the issue and instructed to notify their consignees of the issue and required actions to prevent the occurrence of the anomaly. Please refer to the Customer letter (Ref. CL2018-194_51FS, CL2018-194_3600, CL2018-194_4600, and CL2018-194_5600) for specific details on Required Actions by User. If you have further questions, please contact Ortho Care" Technical Solutions Center at 1-800-421-3311.
Quantity in commerce
1105
Distribution
US Nationwide Distribution
Product codes and lots
Product code: 6801375 UDI: 10758750001132 Serial Numbers: J34000132- J34002322

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0651-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0650-2019Class IITerminatedVoluntary: Firm initiated
Z-0651-2019Class IITerminatedVoluntary: Firm initiated
Z-0652-2019Class IITerminatedVoluntary: Firm initiated
Z-0653-2019Class IITerminatedVoluntary: Firm initiated