UroLift System REF Catalog Number: UL400-4 Tray Contents: UroLift System with Handle Release Tool STERILE R...
Neotract Inc began a recall of UroLift System REF Catalog Number: UL400-4 Tray Contents: UroLift System with Handle Release Tool STERILE R CE 0050 Manufactured By: neotract on . FDA classified it as Class II and gives the reason as "Upon implant deployment, the Capsular Tab may not be delivered as the needle is retracted. In this failure mode, the needle is deployed into the prostate and retracted, leaving no implant behind. Failure to deliver a...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0598-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Neotract Inc
- Product
- UroLift System REF Catalog Number: UL400-4 Tray Contents: UroLift System with Handle Release Tool STERILE R CE 0050 Manufactured By: neotract
- Product code
- PEW Implantable Transprostatic Tissue Retractor System
- Reason for recall
- Upon implant deployment, the Capsular Tab may not be delivered as the needle is retracted. In this failure mode, the needle is deployed into the prostate and retracted, leaving no implant behind. Failure to deliver a capsular tab may result in a delay in completing a treatment or an inability to complete a treatment for the patient. Use of a recalled device may lead to increased frequency of exiting known risks associated with the product, including bleeding and tissue trauma associated with delivering the needle.
- FDA root cause
- Finished device change control
- Action
- On July 6, 2018, the firm sent an Urgent Medical Device Recall Notification to customers via Federal Express next day Service informing them that upon implant deployment, the capsular tab may not be delivered as the needle is retracted. In this failure mode, the needle is deployed into the prostate and retracted, leaving no implant behind. Failure to deliver a capsular tab may result in a delay in completing a treatment or an inability to complete a treatment for the patient. Use of a recalled device may lead to increased frequency of existing known risks associated with the product, including bleeding and tissue trauma associated with delivering the needle. It is important to note that this issue impacts only the delivery device. The implant is not impacted. Any implants that have been delivered with the device are not affected. The notification also provides instructions and the following actions to take: 1) If you HAVE affected stock: a. Immediately discontinue use and quarantine any products with the lot numbers listed above, so that the affected products can be returned to the firm. b. Please return the product as follows: i. Complete the enclosed Recall Acknowledgement Form and email to [email protected] or fax it to (925) 401-0699, Attn: Customer Service. This will allow us to document the amount of product you have on hand for return. A customer service representative will issue you a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to the firm. ii. Write the RGA Number on each box and follow the shipping instructions provided by the firm's Customer Service. 2) If you HAVE NO affected stock: a. Please complete the enclosed Recall Acknowledgement Form and email to [email protected] or fax it to (925) 401-0699, Attn: Customer Service. b. This will allow us to document your receipt of this letter. If customers have any other questions, they can contact their local Urology Consu
- Quantity in commerce
- 8,804
- Distribution
- US: AK, AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV,
- Product codes and lots
- Model: AN00141 (US) and AN00142 (OUS); Catalog No.: UL400-4; Lots: P37940, P37943, P37953, P37956, P37965, P37971, P37985, P37988, P37995, P38001, P38013, P38020, P38124, P38132
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0598-2019 | Class II | Terminated | Voluntary: Firm initiated |