Covidien Emprint Standard Percutaneous Antenna with Thermosphere Technology, CA20L2 intended for use in...
Covidien LLC began a recall of Covidien Emprint Standard Percutaneous Antenna with Thermosphere Technology, CA20L2 intended for use in percutaneous, laparoscopic, and intraoperative... on . FDA classified it as Class II and gives the reason as "The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0627-2019
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic, Inc.
- Product
- Covidien Emprint Standard Percutaneous Antenna with Thermosphere Technology, CA20L2 intended for use in percutaneous, laparoscopic, and intraoperative coagulation (ablation) of soft tissue, including partial or complete ablation of nonresectable liver tumors.
- Product code
- NEY System, Ablation, Microwave And Accessories
- Reason for recall
- The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On November 15, 2018 Medtronic issued UPDATED URGENT MEDICAL DEVICE RECALL notices to customers. Customers are advised to take the following actions: 1. Please immediately quarantine and discontinue use of the affected products. 2. Return affected product as indicated. 3. If you have distributed the products, please promptly forward the information from the letter to those recipients. 4. Complete the Recalled Product Return Form even if you do not have inventory. If you have any questions or concerns, please do not hesitate to contact your Medtronic representative or Technical Support at (800) 448-3644, Option 3, Option 8.
- Distribution
- US Nationwide
- Product codes and lots
- GTIN Number - 10884521706590 Lot Numbers- S7HG001PX S7HG004KX S7HG010KX S7HG011KX S7HG012KX S7HG030LX S7HG031LX S7HG035LX S7HG039LX S7HG042LX S7HG043X S7JG006X S7JG010X S7JG013X S7JG014X S7KG008LX S7KG009LX S7KG010LX S7KG041X S7KG043X S7LG003X S7LG004X S7MG003X S7MG004LX S7MG017PX S7MG029X S8BG001X S8BG002X S8BG006X S8CG001X S8CG008LX S8CG009X S8CG018X S8CG023X S8DG019LX S8DG020LX S8DG021LX S8DG022LX S8DG023X S8DG025X
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0625-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0626-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0627-2019 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
