AIA-PACK B12 (Vitamin B12) is designed for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of...
Tosoh Bioscience Inc began a recall of AIA-PACK B12 (Vitamin B12) is designed for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of vitamin B12 (B12) in human serum on Tosoh AIA... on . FDA classified it as Class II and gives the reason as "Potential for fluorescein interference to cause inaccurately elevated results for certain Tosoh AIA assays". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0671-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Tosoh Bioscience Inc
- Product
- AIA-PACK B12 (Vitamin B12) is designed for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of vitamin B12 (B12) in human serum on Tosoh AIA System analyzers.
- Product code
- CDD Radioassay, Vitamin B12
- Reason for recall
- Potential for fluorescein interference to cause inaccurately elevated results for certain Tosoh AIA assays
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On November 30, 2018, Tosoh Bioscience issued Urgent Medical Device Recall notices to customers. Customers are advised to take the following actions: " Do not use the assays identified in notice for those patients who have recently had a fluorescein angiography procedure. For these patients, use an alternative testing method that does not use a fluorescein-based technology. " Inform medical professionals that the Tosoh AIA assays in the notice utilize a fluorescein-based technology and should not be used to test patient samples for patients who have recently undergone fluorescein angiography. " Continue to use the Tosoh AIA assays listed in notice only for patients who have not undergone a fluorescein angiography procedure exam, or on patient samples which have been taken prior to receiving the fluorescein angiography procedure. " Complete & return the Urgent Medical Device Recall Acknowledgement Form within 15-days of receiving the notification. " Maintain a copy of the notification with your laboratory records and forward this information to others who may have received this product. Should you have any questions, please contact Bernadette OConnell at 1(800) 248-6764 or by email at [email protected]. She will be able to answer any questions Monday-Friday from 9 AM to 5 PM (PST).
- Quantity in commerce
- 29645 units
- Distribution
- US Nationwide
- Product codes and lots
- all lots
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0671-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0672-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0673-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0674-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0675-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0676-2019 | Class II | Terminated | Voluntary: Firm initiated |