Z-0550-2019Class II
Captus 3000 Thyroid Uptake System, Model # 5430-0076 and 5430-0077
Mirion Technologies (Capintec), Inc. / initiated 2018-11-05 / Terminated
Capintec Inc began a recall of Captus 3000 Thyroid Uptake System, Model # 5430-0076 and 5430-0077 on . FDA classified it as Class II and gives the reason as "There is a potential for the spring arm failure as a result of a broken tension rod (a component inside the arm). A break in the tension rod can casue the collimator to fall downward to its lowest point of travel...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0550-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Mirion Technologies (Capintec), Inc.
- Product
- Captus 3000 Thyroid Uptake System, Model # 5430-0076 and 5430-0077
- Product code
- IZD Probe, Uptake, Nuclear
- Reason for recall
- There is a potential for the spring arm failure as a result of a broken tension rod (a component inside the arm). A break in the tension rod can casue the collimator to fall downward to its lowest point of travel, approximately 25 inches from the ground.
- FDA root cause
- Under Investigation by firm
- Action
- Notification letters were sent to customers on 11/5/18. A Revised Product Alert - Corrective Action letter was sent to customers beginning 1/15/19 due to an expansion of the event.
- Quantity in commerce
- 385
- Distribution
- Worldwide distribution - US nationwide distributed in the states of AR, AZ, CA, CO, CT, DC, FL, IA, IN, LA, MD, MI, MN, NC, NJ, OH, OK, PA, SC, SD, TX, VA, WI, WV, and countries of Argentina, Brazil, Canada, China, Cyprus, Greece, Hong Kong, India, Japan, South Korea, Vietnam.
- Product codes and lots
- Serial Numbers 901100 through 901309. Serial Numbers 901310 through 901965 as of 1/15/19.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0550-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0551-2019 | Class II | Terminated | Voluntary: Firm initiated |