Z-0550-2019Class II

Captus 3000 Thyroid Uptake System, Model # 5430-0076 and 5430-0077

Mirion Technologies (Capintec), Inc. / initiated 2018-11-05 / Terminated

Capintec Inc began a recall of Captus 3000 Thyroid Uptake System, Model # 5430-0076 and 5430-0077 on . FDA classified it as Class II and gives the reason as "There is a potential for the spring arm failure as a result of a broken tension rod (a component inside the arm). A break in the tension rod can casue the collimator to fall downward to its lowest point of travel...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0550-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Mirion Technologies (Capintec), Inc.
Product
Captus 3000 Thyroid Uptake System, Model # 5430-0076 and 5430-0077
Product code
IZD Probe, Uptake, Nuclear
Reason for recall
There is a potential for the spring arm failure as a result of a broken tension rod (a component inside the arm). A break in the tension rod can casue the collimator to fall downward to its lowest point of travel, approximately 25 inches from the ground.
FDA root cause
Under Investigation by firm
Action
Notification letters were sent to customers on 11/5/18. A Revised Product Alert - Corrective Action letter was sent to customers beginning 1/15/19 due to an expansion of the event.
Quantity in commerce
385
Distribution
Worldwide distribution - US nationwide distributed in the states of AR, AZ, CA, CO, CT, DC, FL, IA, IN, LA, MD, MI, MN, NC, NJ, OH, OK, PA, SC, SD, TX, VA, WI, WV, and countries of Argentina, Brazil, Canada, China, Cyprus, Greece, Hong Kong, India, Japan, South Korea, Vietnam.
Product codes and lots
Serial Numbers 901100 through 901309. Serial Numbers 901310 through 901965 as of 1/15/19.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0550-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0550-2019Class IITerminatedVoluntary: Firm initiated
Z-0551-2019Class IITerminatedVoluntary: Firm initiated