Z-0551-2019Class II
Captus 700 Thyroid Uptake System, Model # 5430-3137 and 5430-3138
Mirion Technologies (Capintec), Inc. / initiated 2018-11-05 / Terminated
Capintec Inc began a recall of Captus 700 Thyroid Uptake System, Model # 5430-3137 and 5430-3138 on . FDA classified it as Class II and gives the reason as "There is a potential for the spring arm failure as a result of a broken tension rod (a component inside the arm). A break in the tension rod can casue the collimator to fall downward to its lowest point of travel...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0551-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Mirion Technologies (Capintec), Inc.
- Product
- Captus 700 Thyroid Uptake System, Model # 5430-3137 and 5430-3138
- Product code
- IZD Probe, Uptake, Nuclear
- Reason for recall
- There is a potential for the spring arm failure as a result of a broken tension rod (a component inside the arm). A break in the tension rod can casue the collimator to fall downward to its lowest point of travel, approximately 25 inches from the ground.
- FDA root cause
- Under Investigation by firm
- Action
- Notification letters were sent to customers on 11/5/18. A Revised Product Alert - Corrective Action letter was sent to customers beginning 1/15/19 due to an expansion of the event.
- Quantity in commerce
- 65
- Distribution
- Worldwide distribution - US nationwide distributed in the states of AR, AZ, CA, CO, CT, DC, FL, IA, IN, LA, MD, MI, MN, NC, NJ, OH, OK, PA, SC, SD, TX, VA, WI, WV, and countries of Argentina, Brazil, Canada, China, Cyprus, Greece, Hong Kong, India, Japan, South Korea, Vietnam.
- Product codes and lots
- Serial Numbers 700004 through 700018. Serial Numbers 700019 through 700086 as of 1/15/19.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0550-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0551-2019 | Class II | Terminated | Voluntary: Firm initiated |