Alinity i TSH Reagent Kit, List Number 07P4830
Abbott Ireland Diagnostics Division began a recall of Alinity i TSH Reagent Kit, List Number 07P4830 on . FDA classified it as Class II and gives the reason as "Abbott has become aware of a manufacturing issue that can result in damage to a bottle neck in some of the reagent cartridges found in the lots listed in Appendix A. If a bottle neck is damaged, it may cause a failure...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0664-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Ireland Diagnostics Division
- Product
- Alinity i TSH Reagent Kit, List Number 07P4830
- Product code
- JLW Radioimmunoassay, Thyroid-Stimulating Hormone
- Reason for recall
- Abbott has become aware of a manufacturing issue that can result in damage to a bottle neck in some of the reagent cartridges found in the lots listed in Appendix A. If a bottle neck is damaged, it may cause a failure during the reagent loading process on the Alinity i instrument.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The firm initiated the recall by letter on 11/21/2018. The notice explained the problem and provided instruction on how to inspect the reagent cartridges for defect. The consignee was directed to discard the defective units. The units will be replaced.
- Quantity in commerce
- 678 units
- Distribution
- AR, FL, GA, MD, OK, VA
- Product codes and lots
- UDI/Lot Numbers: (01)00380740131159 (17)190710(10)90061UI00 / 90061UI00 (01)00380740131159 (17) 190827(10) 91526UI00 / 91526UI00 (01)00380740131159 (17)190924(10)92291UI00 / 92291UI00
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0664-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0665-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0666-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0667-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0668-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0669-2019 | Class II | Terminated | Voluntary: Firm initiated |