Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0669-2019Class II

Alinity i 25-OH Vitamin D Reagent Kit, List Number 08P4532

Abbott Ireland Diagnostics Division / initiated 2018-11-21 / Terminated / root cause: manufacturing process

Abbott Ireland Diagnostics Division began a recall of Alinity i 25-OH Vitamin D Reagent Kit, List Number 08P4532 on . FDA classified it as Class II and gives the reason as "Abbott has become aware of a manufacturing issue that can result in damage to a bottle neck in some of the reagent cartridges found in the lots listed in Appendix A. If a bottle neck is damaged, it may cause a failure...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0669-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Ireland Diagnostics Division
Product
Alinity i 25-OH Vitamin D Reagent Kit, List Number 08P4532
Product code
MRG System, Test, Vitamin D
Reason for recall
Abbott has become aware of a manufacturing issue that can result in damage to a bottle neck in some of the reagent cartridges found in the lots listed in Appendix A. If a bottle neck is damaged, it may cause a failure during the reagent loading process on the Alinity i instrument.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm initiated the recall by letter on 11/21/2018. The notice explained the problem and provided instruction on how to inspect the reagent cartridges for defect. The consignee was directed to discard the defective units. The units will be replaced.
Quantity in commerce
760 units
Distribution
AR, FL, GA, MD, OK, VA
Product codes and lots
UDI/Lot Numbers (01)00380740121952 (17)191011(10)92313UI00 / 92313UI00  (01)00380740121952 (17)190831(10)92484UI00 / 92484UI00

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0669-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0664-2019Class IITerminatedVoluntary: Firm initiated
Z-0665-2019Class IITerminatedVoluntary: Firm initiated
Z-0666-2019Class IITerminatedVoluntary: Firm initiated
Z-0667-2019Class IITerminatedVoluntary: Firm initiated
Z-0668-2019Class IITerminatedVoluntary: Firm initiated
Z-0669-2019Class IITerminatedVoluntary: Firm initiated