Reliance LITe Nerve Hook Bayoneted, 90, Blunt Tip, REF # 48066103 The blunt probe can be used as a...
Howmedica Osteonics Corp. began a recall of Reliance LITe Nerve Hook Bayoneted, 90, Blunt Tip, REF # 48066103 The blunt probe can be used as a conventional means of achieving nerve retraction. The bend... on . FDA classified it as Class II and gives the reason as "Instead of having a spherical diameter, the probe was observed to have 2 radii and a flat surface. Additionally, upon further investigation, it was observed that some parts exhibited an edge on the intersection between...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0657-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Reliance LITe Nerve Hook Bayoneted, 90, Blunt Tip, REF # 48066103 The blunt probe can be used as a conventional means of achieving nerve retraction. The bend at the proximal tip of the ring the discectomy, as well as during the insertion of an IBD during a conventional TLIF procedure.
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- Instead of having a spherical diameter, the probe was observed to have 2 radii and a flat surface. Additionally, upon further investigation, it was observed that some parts exhibited an edge on the intersection between the radius and the diameter of the probe.
- FDA root cause
- Under Investigation by firm
- Action
- The firm, Stryker, sent an "Urgent Medical Device Recall" letter dated 11/14/2018 to its customers on 11/26/2018. The letter described product, problem and actions to be taken. The customers were instructed to: Examine your inventory; Immediately quarantine the affected product; and complete customer response form to Stryker Spine at email: [email protected]. For questions contact Manager-Regulatory Compliance & Quality System at 866-45-SPINE (77463)..
- Quantity in commerce
- 39
- Distribution
- Worldwide distribution: US Distribution to the state of: PA and countries of:: Australia, Canada, France, Italy, Netherlands, Poland, and Spain.
- Product codes and lots
- Lot #s 183432, 183433, 183434, 17E168
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0657-2019 | Class II | Terminated | Voluntary: Firm initiated |