Protexis Neoprene Surgical Glove, Size 8.5 This powder-free sterile light brown colored surgeons gloves are a...
Cardinal Health 200, LLC began a recall of Protexis Neoprene Surgical Glove, Size 8.5 This powder-free sterile light brown colored surgeons gloves are a disposable device made of synthetic rubber... on . FDA classified it as Class II and gives the reason as "Labeling issue in that the breakthrough (i.e., permeation) times for two chemotherapy drugs (Carmustine and Thiotepa) stated on the wallet (primary packaging) of the Protexis Neoprene surgical gloves are incorrect...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0637-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Protexis Neoprene Surgical Glove, Size 8.5 This powder-free sterile light brown colored surgeons gloves are a disposable device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.
- Product code
- KGO Surgeon'S Gloves
- Reason for recall
- Labeling issue in that the breakthrough (i.e., permeation) times for two chemotherapy drugs (Carmustine and Thiotepa) stated on the wallet (primary packaging) of the Protexis Neoprene surgical gloves are incorrect. Additionally, breakthrough times for Mitomycin were added to the same label in error.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling mix-ups
- Action
- Cardinal Health notified customers on about 10/31/2018 via "URGENT: PRODUCT CORRECTION NOTICE." Customers were instructed to examine inventory for affected products, affix the provided warning label on the principal display panel on the affected dispenser packs, notify internal parties and/or customers if the product was further distributed of the recall, and complete and return the provided Acknowledgement Form. If the wallets containing the gloves are removed from the dispenser packs, customers were also instructed to provide a warning label to accompany the affected product inform users of the correct breakthrough times. If customers need additional labels, they were instructed to contact 800-292-9332 or [email protected] for assistance. For further questions regarding the Product Correction Notice, please reference the following Customer Service telephone numbers: Hospital (800-964-5227), Federal Government (800-444-1166), Distributor (800-635-6021), All other customers (888-444-5440).
- Quantity in commerce
- 1,475,095 pairs total
- Distribution
- Worldwide Distribution: US (Nationwide distribution) to states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and to countries of: Austria, Belgium, Canada, Chile, Costa Rica, France, Germany, Italy, Luxembourg, Martinique, Monaco, Netherlands, Poland, Reunion, Spain, Sweden, Switzerland, Taiwan, and United Kingdom.
- Product codes and lots
- Product code: 2D73DP85, Lot Numbers: TS18060015, TS18060329, TS18070121, TS18070314, TS18080034, TS18080352, and TS18090219.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0631-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0632-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0633-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0634-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0635-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0636-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0637-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0638-2019 | Class II | Terminated | Voluntary: Firm initiated |