Discovery NM/CT 670 DR, model 5376204-70-56 intended for use in detecting, localizing, diagnosing of diseases...
GE Healthcare, LLC began a recall of Discovery NM/CT 670 DR, model 5376204-70-56 intended for use in detecting, localizing, diagnosing of diseases and organ function for the evaluation of... on . FDA classified it as Class II and gives the reason as "When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off...". As of , the recall status on record is Completed.
Recall record
- Recall number
- Z-0606-2019
- Classification
- Class II
- Status
- Completed
- Initiated
- Posted
- Recalling firm
- GE Healthcare
- Product
- Discovery NM/CT 670 DR, model 5376204-70-56 intended for use in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer.
- Product code
- KPS System, Tomography, Computed, Emission
- Reason for recall
- When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off all power to the entire system as intended, resulting in a potential electric hazard.
- FDA root cause
- Component change control
- Action
- The firm initiated the recall by letter on 05/24/2018. The letter explained the issue and requested the UPS(s) be disconnected from the system until the system could be repaired.
- Quantity in commerce
- 4 units
- Distribution
- CA, DC, OH, UT, WI, Canada
- Product codes and lots
- GTIN 00840682124393 Serial Numbers: DDRW34032, DDRX34038, DDRX34041, DDRX34045
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0606-2019 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-0607-2019 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-0608-2019 | Class II | Completed | Voluntary: Firm initiated |