Z-0607-2019Class II

Discovery NM/CT 670 ES, model 5376204-70-57

GE Healthcare / initiated 2018-05-24 / Completed

GE Healthcare, LLC began a recall of Discovery NM/CT 670 ES, model 5376204-70-57 on . FDA classified it as Class II and gives the reason as "When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off...". As of , the recall status on record is Completed.

Recall record

Recall number
Z-0607-2019
Classification
Class II
Status
Completed
Initiated
Posted
Recalling firm
GE Healthcare
Product
Discovery NM/CT 670 ES, model 5376204-70-57
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off all power to the entire system as intended, resulting in a potential electric hazard.
FDA root cause
Component change control
Action
The firm initiated the recall by letter on 05/24/2018. The letter explained the issue and requested the UPS(s) be disconnected from the system until the system could be repaired.
Quantity in commerce
2 units
Distribution
CA, DC, OH, UT, WI, Canada
Product codes and lots
GTIN 00840682123877 Serial Numbers: ESDX35013, ESDW35010

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0607-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0606-2019Class IICompletedVoluntary: Firm initiated
Z-0607-2019Class IICompletedVoluntary: Firm initiated
Z-0608-2019Class IICompletedVoluntary: Firm initiated