Z-0608-2019Class II
Discovery NM/CT 670 Pro, model 5376204-70-54
GE Healthcare / initiated 2018-05-24 / Completed
GE Healthcare, LLC began a recall of Discovery NM/CT 670 Pro, model 5376204-70-54 on . FDA classified it as Class II and gives the reason as "When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off...". As of , the recall status on record is Completed.
Recall record
- Recall number
- Z-0608-2019
- Classification
- Class II
- Status
- Completed
- Initiated
- Posted
- Recalling firm
- GE Healthcare
- Product
- Discovery NM/CT 670 Pro, model 5376204-70-54
- Product code
- KPS System, Tomography, Computed, Emission
- Reason for recall
- When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off all power to the entire system as intended, resulting in a potential electric hazard.
- FDA root cause
- Component change control
- Action
- The firm initiated the recall by letter on 05/24/2018. The letter explained the issue and requested the UPS(s) be disconnected from the system until the system could be repaired.
- Quantity in commerce
- 1 unit
- Distribution
- CA, DC, OH, UT, WI, Canada
- Product codes and lots
- GTIN 00840682121194 Serial Number ESDX35015
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0606-2019 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-0607-2019 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-0608-2019 | Class II | Completed | Voluntary: Firm initiated |