VITROS Anti-HBe Reagent Pack, Cat No. 6801816
ORTHO-CLINICAL DIAGNOSTICS began a recall of VITROS Anti-HBe Reagent Pack, Cat No. 6801816 Product Usage: For the in vitro qualitative detection of antibodies to hepatitis B e antigen (anti-HBe) in human... on . FDA classified it as Class II and gives the reason as "The firm is extending their previous July 2018 recall to include additional product codes. Biased results may occur for these additional products at biotin concentrations which are lower than indicated in the current...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0687-2019
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product
- VITROS Anti-HBe Reagent Pack, Cat No. 6801816 Product Usage: For the in vitro qualitative detection of antibodies to hepatitis B e antigen (anti-HBe) in human adult and pediatric (2 to 21 years old) serum from individuals who have symptoms of chronic hepatitis and those who have recovered from HBV infection, using the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System. Further assessment of HBV infection (biochemical, serological and/or nucleic acid testing) is required to define the specific disease state. VITROS Anti-HBe test performance has not been established for the monitoring of HBV disease or therapy. Test performance characteristics have not been established in patients under the age of 2, or in populations of immunocompromised or immunosuppressed patients. This test has not been FDA licensed for the screening of blood, plasma and tissue donors.
- Product code
- LOM Test, Hepatitis B (B Antigen, B Surface Antigen, Be Antigen)
- Reason for recall
- The firm is extending their previous July 2018 recall to include additional product codes. Biased results may occur for these additional products at biotin concentrations which are lower than indicated in the current Instructions for Use.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On November 1, 2018, a customer letter was sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail to all customers who could process the affected products to inform them of the potential interference of biotin at levels lower than stated in the IFUs and that Ortho is in the process of revising the IFUs for the affected products. Ortho recommended customers follow normal laboratory procedures for troubleshooting of samples containing other assay interferences and it was acceptable to continue using the affected products. If you have questions, please contact Ortho Care Technical Solutions Center at 1-800-421-3311.
- Quantity in commerce
- 3429 units (100 tests per unit)
- Distribution
- Worldwide Distribution - US Nationwide domestic distribution. International distribution to Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, India, France, Japan, Germany, Italy, Mexico, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, and United Kingdom.
- Product codes and lots
- All expired, in-date, and future lots released.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0687-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0688-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0689-2019 | Class II | Ongoing | Voluntary: Firm initiated |