Z-0679-2019Class II

RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayStation 7 and...

RaySearch Laboratories AB (publ) / initiated 2018-11-12 / Terminated

RAYSEARCH LABORATORIES AB began a recall of RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayStation 7 and RayStation 8A on . FDA classified it as Class II and gives the reason as "Isocenter shifts when using dose tracking in RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayStation 7 and RayStation 8A. In these versions, isocenter shifts in dose...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0679-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
RaySearch Laboratories AB (publ)
Product
RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayStation 7 and RayStation 8A
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
Isocenter shifts when using dose tracking in RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayStation 7 and RayStation 8A. In these versions, isocenter shifts in dose tracking are incorrectly interpreted according to the DICOM patient coordinate system.
FDA root cause
Under Investigation by firm
Action
RayStation issued Field Safety Notice, Medical Device Correction #37483, on 11/12/18 via email, states issue, action to be taken and correction of software in Dec 2018.
Quantity in commerce
69
Distribution
AZ CA CO CT DE FL GA HI ID IL LA MA MD ME MI MO MS NC NJ NY OH OR PA RI TN TX VA VT WA WI
Product codes and lots
Software version number:  3.5.0.16 , 3.5.1.6 , 4.0.0.14 , 4.0.1.4 , 4.0.2.9 , 4.0.3.4 , 4.5.0.19 , 4.5.1.14 , 4.5.2.7 , 4.7.0.15 , 4.7.1.10 , 4.7.2.5 , 4.7.3.13 , 4.7.4.4 , 4.7.5.4 , 5.0.0.37 , 5.0.1.11 , 5.0.2.35 , 6.0.0.24 , 6.1.0.26, 6.1.1.2 , 6.2.0.7 , 7.0.0.19 , 8.0.0.61 or 8.0.1.10

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0679-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0679-2019Class IITerminatedVoluntary: Firm initiated